zeiss_axiovert_200m

This is an old revision of the document!


Location: Room P2-A-22 ( 47217)
Manufacturer: Carl Zeiss MicroImaging
Model: Axiovert 200M
Nickname: "Axiovert"
Software: Metamorph 7.7.9.0
Year: 2003
SN: ZCA522000494
 
 
 
Zeiss Axiovert 200M Booking
Zeiss Axiovert 200M Quality Control
Zeiss Axiovert 200M Usage Statistics

Microscope overview

The Zeiss Axiovert 200M is a fully motorized inverted widefield fluorescence microscope ideal for live-cell imaging applications. It is equipped with a small stage incubator for temperature control and CO2 supply. Using the Metamorph software you can control the motorized filters, shutters and stage to simultaneously set up multi-color time-lapse imaging experiments in multiple stage positions. Its sensitive cooled CCD camera (Photometrics Coolsnap HQ CCD) allows you to capture weak fluorescent signals and minimize photobleaching/photodamage in light sensitive samples. You can also perform Z-stack acquisition and use PPBI Huygens RM deconvolution server to significantly improve image resolution even in thick samples. If you are performing long time-lapse experiments in live samples (aqueous media) and need hardware focus control, check the Zeiss Cell Observer with Definite Focus. If don't want to use deconvolution, check out one of the confocal systems such as the point scanners Zeiss LSM 710 and Zeiss LSM 880 or the spinning disks 3i Marianas SDC and Zeiss Cell Observer SD. If your personal computer cannot handle all the data you collected, check out the Big Guy or Colossus. If you have FISH or other fixed fluorescence slides try the Leica DM5000B instead.

Data files older than 1 month will be automatically deleted on this system, please copy your data to the iMM server using the desktop link.

Booking Rules

The Axiovert 200M booking is divided in three daily slots:

9:00 to 13:00 13:00 to 17:00 17:00 to 21:00
  1. A user can book at maximum 2 sessions per week
  2. Each session must fit inside one of the three daily slots (i.e. a session booked from 11:00 to 14:00 counts as 2 sessions)
  3. The above rules do not apply for weekends and for work days before 9:00 and after 21:00
  4. The above rules do not apply for sessions booked on the very same day or 24 hours ahead
  5. Exceptions to these rules require approval from José Rino.

System components

Filtersets

Position Filterset Reference Excitation Emission
1 Blue FS01 359-371 nm > 397 nm
2 Green BP FS10 450-490 nm 515-565 nm
3 Red FS20 540-552 nm 575-640 nm
4 Far Red Cy5-4040C-000 608-648 nm 672-712 nm
5 A none none none

Upon request:

Position Filterset Reference Excitation Emission
2 Green LP FS09 450-490 nm > 515 nm

Objectives

Magnification Model Immersion NA WD (mm) pixel size* (μm) 10 μm = (color) 10 μm = Reference
10x EC Plan-Neofluar Air 0.30 5.2 1.00 10 pixels 23.3 pixels 420340-9901-000
20x EC Plan-Neofluar Air 0.50 2 0.48 21 pixels 46.7 pixels 420350-9900-000
40x EC Plan-NeoFluar Air 0.75 0.71 0.24 41 pixels 93.7 pixels 440350-9903-000
63x Plan-Apochromat DIC Oil 1.40 0.19 0.15 65 pixels - 420782-9900-000

Upon request:

Magnification Model Immersion NA WD (mm) pixel size* (μm) 10 μm = Reference
1.25x EC Epiplan-Neofluar Air 0.035 4 7.58 1.3 pixels 422310-9900-000
5x EC Plan-NeoFluar Air 0.16 18.5 2.00 5 pixels 420330-9901-000
20x EC Plan-Neofluar Ph 2 Air 0.50 2 0.48 21 pixels 440341-9904-000
40x Plan-Neofluar Ph3 Oil 1.30 0.20 0.24 41 pixels 440451-9903-000
63x Plan-Apochromat Ph3 Oil 1.4 0.19 0.15 65 pixels 420781-9910-000

* pixel sizes for binning = 1

Stage

For automated acquisition of 96 well plates use: X = -9014 μm Y = -8995 μm

System Turn On Procedures

  • Turn the fluorescence light power source on.

  • Turn the Prior microscope stage controller on.

  • Wait 10s then turn the microscope on.

  • Turn the camera on.

  • Turn the computer on.
  • Login to the computer (Axiovert 200M user).
  • Start the Metamorph software.

Microscope Turn Off procedures

  • Turn off the computer.
  • Turn off the camera.
  • If you used any immersion oil in the objectives, clean them up.
  • Turn off the microscope stage controller.
  • Turn off the shutter controller.
  • Turn off the microscope.
  • If there's another user for this microscope in the next hour leave the mercury lamp power source on, else turn it off.
  • Be extra careful in weekends and at the end of the day to make sure that the mercury lamp is not going to be turned on for an excessive time.

Back to the Resources page