Users must get trained on a cell analyzer before using it. Follow the
New User guidelines to start using our equipment.
The cell analyzers can be booked up to 1 month in advance.
Users must include 15 min in their booking time for cleaning.
Users must have a valid reservation in a system before using it.
Users must use their own booked time and not book or use a resource with other username.
Users can only cancel bookings 3h before the slot starting time.
Users must perform an additional cleaning before acquisition if they are the first user of the day.
Users must record start and finish times on the system log sheet (cleaning time included) and problems (if any).
Users that swap filters must leave the system with the default configuration.
Users must always clean the machines after acquisition following the cleaning protocol and record the cleaning steps in the Software.
Users must refill the sheath tank with PBS+Azide at the end of the booked session.
If during acquisition, the waste container gets full, a high pitch sound will go off. Users should disconnect the black sensor (the noise will stop), put the cytometer in Standby, and disconnect
the orange connector. The container must be replaced with a new one.
Users must always export their data and delete it from the Software database regularly.
Users must log out from the analyzer software at the end of their session.
(BD LSRFortessa, LSRFortessa SORP and Fortessa X-20) Users must switch off the system if they are the last users booked or if the analyzer will not be used for at least 3h.
(BD Accuri C6 Plus, BD FACSymphony A5 SE and Aurora) After 5pm or on weekends and holidays, users must switch off the system if they are the last users booked.