User Tools

Site Tools


zeiss_lsm_880

This is an old revision of the document!


Location: [Lisbon] Room P2-A-22 ( 47217)
Manufacturer: ZEISS Microscopy
Model: LSM 880 with Airyscan (click for Zeiss webinar)
Nickname: "880"
Software: ZEN 2.3 SP1 (black edition)
Year: 2016
SN: 2811000267
 
Data will be deleted after: 1 month  
 
Zeiss LSM 880 Quality Control
Zeiss LSM 880 Usage Statistics

Microscope overview

The Zeiss LSM 880 with Airyscan is a point scanning confocal microscope able to generate high-resolution three-dimensional images of thick specimens with high sensitivity, high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments, equipped with a large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions. Its Airyscan with 32 concentrically arranged GaAsP detection elements is able to collect all light from an Airy pattern simultaneously, allowing for increased resolution (140 nm laterally and 400 nm axially, at 488 nm). It can also be used in Fast Mode, which enables parallel excitation and detection of four image pixels. The result is a speed improvement by a factor 4x, while maintaining the sensitivity of Airyscan and 1.5x resolution improvement. Its scanning unit further includes two PMT detectors (one of them cooled) as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 458, 488, 514, 561, 594 and 633 nm excitation wavelengths).

Data files older than 1 month will be automatically deleted on this system, please copy your data to the GIMM server using the desktop link.

System components

LASERs

Laser Unit Wavelength Maximum Power
violet 405 nm 30 mW
Argon 458, 488 and 514 nm 25 mW
yellow-green 561 nm 20 mW
orange 594 nm 2 mW
red 639 nm 5 mW

Objectives

Magnification Model Immersion NA WD (mm) Reference
10x EC Plan-Neofluar Air 0.30 5.20 420340-9901-000
20x Plan-Apochromat Air 0.80 0.55 420650-9901-000
40x C-Apochromat Corr Water 1.20 0.28 421767-9970-000
40x LD C-Apochromat Corr Water 1.10 0.62 421867-9970-000
63x Plan-Apochromat DIC Oil 1.40 0.19 420782-9900-799

Upon request:

Magnification Model Immersion NA WD (mm) Reference
25x LCI Plan-Neofluar Corr DIC Oil/Glyc/W 0.80 0.21 420852-9972-000
40x Plan-Apochromat Corr Dry 0.95 0.25 420660-9970-000

Filtersets (for eyepieces)

Position Filterset Reference Excitation Dichroic Emission
2 Cyan 49001 426-446 nm 455 nm 460-500 nm
3 Green FS38HE 450-490 nm 493 nm 500-550 nm
4 Red FS43 533-558 nm 570 nm 570-640 nm
5 Blue FS49 G 365 nm 395 nm 420-470 nm

How to start your session

If you are the first user of the day:

zeiss_lsm_880_1.jpg

  • Turn off Components
  • Turn off System/PC
  • Turn off Main Switch
  • Turn on the computer
  • Turn on Main Switch
  • Turn on Systems/PC
  • Turn on Components
  • Turn on HXP Lamp
  • Log into Windows with your Agendo credentials
  • Start ZEN Black

Else:

  • Log into Windows with your Agendo credentials
  • Start ZEN Black

How to end your session

If there is another user:

  • Clean up immersion objectives (gently wipe with optic cleaning paper and EtOH embedded paper)
  • Exit the ZEN software
  • Log off the computer

If you are the last user of the day:

  • Clean up immersion objectives (gently wipe with optic cleaning paper and EtOH embedded paper)
  • Turn off LASERs on ZEN
  • Exit the ZEN software
  • Turn off HXP lamp
  • Turn off the computer

Back to the Equipment page

zeiss_lsm_880.1765230897.txt.gz ยท Last modified: 2025/12/08 22:54 by bioimaging