zeiss_lsm_880

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
zeiss_lsm_880 [2020/06/01 19:22]
bioimaging
zeiss_lsm_880 [2020/12/02 11:29]
bioimaging
Line 8: Line 8:
 <​b>​Location</​b>:​ <a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=zeiss_lsm_880_map_new.png">​Room P2-A-22</​a>​ (<img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=phone_neg.png"​ width=15>​ 47217) <​br> ​ <​b>​Location</​b>:​ <a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=zeiss_lsm_880_map_new.png">​Room P2-A-22</​a>​ (<img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=phone_neg.png"​ width=15>​ 47217) <​br> ​
 <​b>​Manufacturer</​b>:​ <a href="​http://​www.zeiss.de/​micro"​ target="​_blank">​Carl Zeiss MicroImaging</​a><​br>​ <​b>​Manufacturer</​b>:​ <a href="​http://​www.zeiss.de/​micro"​ target="​_blank">​Carl Zeiss MicroImaging</​a><​br>​
-<​b>​Model</​b>:​ LSM 880 with Airyscan<​br>​+<​b>​Model</​b>: ​<a href="​https://​www.youtube.com/​watch?​v=sGTNBXq8omg">​LSM 880 with Airyscan</a> (click for Zeiss webinar) ​<br>
 <​b>​Nickname</​b>:​ "​880"<​br>​ <​b>​Nickname</​b>:​ "​880"<​br>​
-<​b>​Software</​b>:​ ZEN 2.(black) <br>+<​b>​Software</​b>:​ ZEN 2.3 SP1 (black ​edition) <br>
 <​b>​Year</​b>:​ 2016 <br> <​b>​Year</​b>:​ 2016 <br>
 <​b>​SN</​b>:​ 2811000267 <br> <​b>​SN</​b>:​ 2811000267 <br>
Line 29: Line 29:
 ===== Microscope overview ===== ===== Microscope overview =====
  
-{{  :​airyscan_beampath.png?​400}} The Zeiss LSM 880 with Airyscan is a //confocal point-scanning microscope// able to generate high-resolution three-dimensional images of thick specimens with high sensitivity,​ high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments,​ equipped with a large size incubator for temperature control and CO2 supply and with Definite Focus for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions. Its [[http://​www.zeiss.com/​content/​dam/​Microscopy/​Products/​laser-scan/​LSM%20880/​zeiss-lsm880-airyscan-principle-672.jpg|Airyscan]] 32 channel area detector is able to collect all light from an Airy pattern simultaneously,​ allowing for increased resolution (140 nm laterally and 400 nm axially, at 488 nm). It can also be used in [[https://​www.youtube.com/​watch?​v=ln6VUmLzLLc|Fast Mode]], which enables parallel excitation and detection of four image pixels. The result is a speed improvement by a factor of four, while maintaining the sensitivity of Airyscan and 1.5x resolution improvement. Its scanning unit further includes two PMT detectors (one of them cooled) as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 458, 488, 514, 561, 594 and 633 nm excitation wavelengths). ​+{{  :​airyscan_beampath.png?​400}} The Zeiss LSM 880 with Airyscan is a [[https://imm.medicina.ulisboa.pt/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=confocal.png|confocal point-scanning microscope]] able to generate high-resolution three-dimensional images of thick specimens with high sensitivity,​ high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments,​ equipped with a large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions. Its [[http://​www.zeiss.com/​content/​dam/​Microscopy/​Products/​laser-scan/​LSM%20880/​zeiss-lsm880-airyscan-principle-672.jpg|Airyscan]] 32 channel area detector is able to collect all light from an Airy pattern simultaneously,​ allowing for increased resolution (140 nm laterally and 400 nm axially, at 488 nm). It can also be used in [[https://​www.youtube.com/​watch?​v=ln6VUmLzLLc|Fast Mode]], which enables parallel excitation and detection of four image pixels. The result is a speed improvement by a factor of four, while maintaining the sensitivity of Airyscan and 1.5x resolution improvement. Its scanning unit further includes two PMT detectors (one of them cooled) as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 458, 488, 514, 561, 594 and 633 nm excitation wavelengths). ​
 {{::​fast_airyscan.png?​420 ​ }}With this system you can perform optical sectioning of fluorescent samples which are too thick for a widefield system such as the [[zeiss_cell_observer|Zeiss Cell Observer]] or the [[zeiss_axiovert_200m|Zeiss Axiovert 200M]]. Image resolution with Airyscan is higher than the [[zeiss_lsm_710|Zeiss LSM 710]] and detection sensitivity is also higher than the spinning disks [[3i_marianas_sdc|3i Marianas SDC]] and [[zeiss_cell_observer_sd|Zeiss Cell Observer SD]]. If your personal computer cannot handle all the data you collected, check out the [[big_guy|Big Guy]] or [[colossus|Colossus]]. {{::​fast_airyscan.png?​420 ​ }}With this system you can perform optical sectioning of fluorescent samples which are too thick for a widefield system such as the [[zeiss_cell_observer|Zeiss Cell Observer]] or the [[zeiss_axiovert_200m|Zeiss Axiovert 200M]]. Image resolution with Airyscan is higher than the [[zeiss_lsm_710|Zeiss LSM 710]] and detection sensitivity is also higher than the spinning disks [[3i_marianas_sdc|3i Marianas SDC]] and [[zeiss_cell_observer_sd|Zeiss Cell Observer SD]]. If your personal computer cannot handle all the data you collected, check out the [[big_guy|Big Guy]] or [[colossus|Colossus]].
  
Line 36: Line 36:
 {{:​warning.png?​25|}} Data files older than **1 month** will be automatically deleted on this system, please copy your data to the iMM server using the desktop link. {{:​warning.png?​25|}} Data files older than **1 month** will be automatically deleted on this system, please copy your data to the iMM server using the desktop link.
  
-===== Booking Rules ===== +===== Additional information ​=====
-The Zeiss LSM 880 booking is divided in three daily slots: +
-|  9:00 to 13:00  |  13:00 to 17:00  |  17:00 to 21:00  |+
  
-  - **user** can book at maximum **2 sessions ​per week** +  ​* [[https://​www.youtube.com/​watch?​v=sGTNBXq8omg|ZEISS Webinar: LSM 880 with Airyscan ​Revolutionize Your Confocal Imaging (video)]] 
-  - Each **session must fit inside one of the three daily slots** (i.e. a session booked from 11:00 to 14:00 counts as 2 sessions) +  ​[[https://​www.youtube.com/​watch?​v=VuxgaKDa_DU|ZEISS Webinar: Fast Mode for ZEISS LSM 880 with Airyscan (video)]] 
-  - The above rules do not apply for **weekends** and for work days **before 9:00** and **after 21:00** + 
-  - The above rules do not apply for sessions booked on the very same day or **24 hours ahead**+===== Booking Rules ===== 
 +  - **Users** can book at maximum **8 hours per week** 
 +  - This usage restriction does not apply for **weekends** and for working ​days **before 9:00** and **after 21:00**
   - Exceptions to these rules require **approval from [[joserino@medicina.ulisboa.pt|José Rino]]**.   - Exceptions to these rules require **approval from [[joserino@medicina.ulisboa.pt|José Rino]]**.
-  - Users must indicate which lasers they are going to use in the booking Description field 
-  - Users must indicate if they will use the incubator in the booking Description field (write down 37C INCUBATOR) 
  
 ===== System components ===== ===== System components =====
Line 81: Line 79:
 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
   * Turn on the ''​MAIN SWITCH''​   * Turn on the ''​MAIN SWITCH''​
-  * Turn on the two switches: ​''​SYSTEMS/​PC'' ​and ''​COMPONENTS''​+  * Turn on the ''​SYSTEMS/​PC'' ​switch
 {{zeiss_lsm_880_1.jpg?​100}} {{zeiss_lsm_880_1.jpg?​100}}
   * Turn on the computer   * Turn on the computer
   * Login in to Windows (Bioimaging User)   * Login in to Windows (Bioimaging User)
 +  * Turn on the ''​COMPONENTS''​ switch
 +  * Check that the computer is connected to an Unidentified network (click on the network icon in the taskbar)
   * Start the **ZEN Black** software   * Start the **ZEN Black** software
 +
 +**Warning**:​ If you need to change the stage adapter, please contact the <​html><​a href="/​facility/​bioimaging/​doku.php?​id=about">​Bioimaging Unit</​a>​ (<​b>​imm-bioimaging@medicina.ulisboa.pt</​b>​ | <img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=phone_neg.png"​ width=15>​ 47222</​html>​)
  
 ===== Microscope Turn Off Procedure ===== ===== Microscope Turn Off Procedure =====
zeiss_lsm_880.txt · Last modified: 2024/01/28 11:47 by bioimaging