Location: [Oeiras] 0B06
Manufacturer: Prairie (now Bruker!)
Model: Prairie Ultima
Nickname: "Prairie"
Software: PrairieView 4.5
Year: 2008
SN: N/A
 
Data will be deleted after: 3 months  
 
XXX Quality Control
XXX Usage Statistics

Microscope overview

A multi-photon, or 2-photon, microscope uses modules similar to those of a laser scanning confocal microscope: A laser source for sample excitation, a scanhead with galvo mirrors to scan the excitation beam, and photo-multiplier tubes to detect fluorescence signals. 2-photon imaging has advantages over confocal: the use of a near-infrared laser allows deeper penetration into tissues, the radiation is less energetic and potentially less phototoxic, and because the excitation effect is confined to the focal plane, de-scanning and pinholes are not necessary, so there are less light looses on the detection side. This Prairie Ultima system is integrated with a custom modified fixed stage ready for intra-vital imaging, and 2photon is ideal for studies in thick tissue slices, in-vivo, or imaging of small organisms.

Click on the image on the right to see it in higher detail

Data files older than 3 months will be automatically deleted on this system, please copy your data to the institutional server using the desktop link.

System components

LASERs

Laser Unit Wavelength
Coherent - Chameleon 650 - 1050 nm

Objectives

Magnification Model Immersion NA WD (mm) Reference
20x Olympus (Evident now!) XLUM FL Plan Water (dipping) 1.00 2.00 https://www.thorlabs.com/thorproduct.cfm?partnumber=N20X-PFH

Filtersets (Ocular)

Filterset Reference Excitation Dichroic Emission Fluorochromes
Green - 525/50 nm blue > 515 nm GFP, FITC, Alexa488
Orange - 5xx/xx nm green > xxx nm TRITC, Alexa568
Blue - 460x/xx nm violet > xxx nm DAPI

Emission Filters (multi-photon)




Detector Filter Transmission Fluorochromes
CH1 (PMT) Red 660/40 nm Alexa647, Cy5
Ch2 (GaAsP) Green 595/50 nm mCherry, Alexa568, Cy3
Ch3 (GaAsP) Far Red 525/50 nm GFP, Alexa488, FITC
Ch4 (PMT1) Blue 460/50 nm DAPI, CFP or BS 2nd Harmonic




Upon request:
(requires changing filters and the dichroic beamsplitter)




Detector Filter Transmission Fluorochromes
Ch1 - - nm Cy5
Ch2 - - nm Alexa568
Ch3 - - nm Alexa488
Ch4 - - nm DAPI
DODT - - nm BF




UNDER CONSTRUCTION!!!

System Turn On Procedures

  • 1. Check that the cooling and power supply for the laser are ON

  • 2. Turn on the fluorescence lamp

* 3. …

System turn Off procedures

If there is another user for this microscope in the next hour:

  • Leave the fluorescent lamp and lasers on
  • Clean up immersion objectives

Else:

  • Clean up objective(s)
  • Set the laser to standby

Back to the Equipment page