Line 88: Line 88:
  
 The system should be always on, you only need to: The system should be always on, you only need to:
-  * Turn on the computer (if needed)+  ​* If incubation is needed (turn on 30 mins before acquiring):​ 
 +    ​* Turn the powerbar off 
 +{{::​Oeiras_LSM900_powerbar.png?​300|}} 
 +    * Turn the incubation controller ​on 
 +{{::​Oeiras_LSM900_incubation.png?​300|}} 
 +    * Turn the powerbar on
   * Log into Windows with your Agendo credentials   * Log into Windows with your Agendo credentials
   * Start **ZEN**   * Start **ZEN**
Line 113: Line 118:
 {{::​oeiras_900_turningon2.jpg?​300|}} {{::​oeiras_900_turningon2.jpg?​300|}}
   * Turn on the ''​Power bar''​ (on the back right-hand side of the microscope)   * Turn on the ''​Power bar''​ (on the back right-hand side of the microscope)
-{{::oeiras_900_turningon3.jpg?300|}}+{{::oeiras_lsm900_powerbar.png?300|}}
   * Turn on the computer   * Turn on the computer
   * Log into Windows with your Agendo credentials   * Log into Windows with your Agendo credentials
Line 128: Line 133:
   * Turn off the ''​SYSTEM''​ and ''​COMPONENTS''​ switches   * Turn off the ''​SYSTEM''​ and ''​COMPONENTS''​ switches
   * Turn off the Laser key   * Turn off the Laser key
-  * Turn off the ''​MAIN SWITCH''​ 
  
 ---- ----
  