Line 41: Line 41:
  
 ^  Position ​ ^  Filterset ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^  Reference ​ ^ ^  Position ​ ^  Filterset ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^  Reference ​ ^
-|  1  |  Blue  |  359-371 nm  |  395 nm  |  <fc #​be00cc>>​ 397 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​?​s=163176628fcf844&​l=en&p=us&​f=f&​a=v&​b=f&​id=488001-9901-000&o=|FS01]] ​ | +|  1  |  Blue  |  359-371 nm  |  395 nm  |  <fc #​be00cc>>​ 397 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​en/us/​fluorescenceAssistant/​filtersets/​488001-9901-400|FS01]] ​ | 
-|  2  |  Green BP  |  450-490 nm  |  510 nm  |  <fc #​81ff00>​515-565 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​?​s=163176628fcf844&​l=en&p=us&​f=f&​a=v&​b=f&​id=488010-9901-000&o=|FS10]] ​ | +|  2  |  Green BP  |  450-490 nm  |  510 nm  |  <fc #​81ff00>​515-565 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​en/us/​fluorescenceAssistant/​filtersets/​488010-9901-400|FS10]] ​ | 
-|  3  |  Red  |  540-552 nm  |  560 nm  |  <fc #​ffa600>​575-640 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​?​s=163176628fcf844&​l=en&p=us&​f=f&​a=v&​b=f&​id=488020-9901-000&o=|FS20]] ​ | +|  3  |  Red  |  540-552 nm  |  560 nm  |  <fc #​ffa600>​575-640 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​en/us/​fluorescenceAssistant/​filtersets/​488020-9901-400|FS20]] ​ | 
-|  4  |  Far Red  |  608-648 nm  |  660 nm  |  <fc #​ff0000>​672-712 nm</​fc> ​ |  [[http://www.semrock.com/SetDetails.aspx?id=2983|Cy5-4040C-000]] ​ |+|  4  |  Far Red  |  608-648 nm  |  660 nm  |  <fc #​ff0000>​672-712 nm</​fc> ​ |  [[https://www.idex-hs.com/store/​product-detail/​cy5-4040c/​s-000701?prod=400773|Cy5-4040C-000]] ​ |
 |  5  |  none  ||||| |  5  |  none  |||||
  
Line 50: Line 50:
  
 ^  Position ​ ^  Filterset ​ ^  Reference ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^ ^  Position ​ ^  Filterset ​ ^  Reference ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^
-|  2  |  Green LP  |  450-490 nm  |  510 nm  |  <fc #​1fff00>>​ 515 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​?​s=163176628fcf844&​l=en&p=us&​f=f&​a=v&​b=f&​id=488009-9901-000&o=|FS09]] ​ |+|  2  |  Green LP  |  450-490 nm  |  510 nm  |  <fc #​1fff00>>​ 515 nm</​fc> ​ |  [[https://​www.micro-shop.zeiss.com/​en/us/​fluorescenceAssistant/​filtersets/​488009-9100-400|FS09]] ​ |
  
 ==== Objectives ==== ==== Objectives ====
 ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  pixel size* (μm)  ^  10 μm =  ^ (color) ​ 10 μm =  ^  Reference ​ ^ ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  pixel size* (μm)  ^  10 μm =  ^ (color) ​ 10 μm =  ^  Reference ​ ^
-|  10x  |  EC Plan-Neofluar ​ |  Air  |  0.30  |  5.20  |  1.00  |  10 pixels ​ |  23.3 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420340-9900-000/​Objective-EC-Plan-Neofluar-10x-0.30-M27|420340-9900-000]] ​ | +|  ​<fc #fef042>10x</​fc> ​ ​| ​ EC Plan-Neofluar ​ |  Air  |  0.30  |  5.20  |  1.00  |  10 pixels ​ |  23.3 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420340-9900-000/​Objective-EC-Plan-Neofluar-10x-0.30-M27|420340-9900-000]] ​ | 
-|  20x  |  EC Plan-Neofluar ​ |  Air  |  0.50  |  2.00  |  0.48  |  21 pixels ​ |  46.7 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440340-9904-000/​Objective-EC-Plan-Neofluar-20x-0.5|440340-9904-000]] ​ | +|  ​<fc #0aa05c>20x</​fc> ​ ​| ​ EC Plan-Neofluar ​ |  Air  |  0.50  |  2.00  |  0.48  |  21 pixels ​ |  46.7 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440340-9904-000/​Objective-EC-Plan-Neofluar-20x-0.5|440340-9904-000]] ​ | 
-|  40x  |  EC Plan-NeoFluar ​ |  Air  |  0.75  |  0.71  |  0.24  |  41 pixels ​ |  93.7 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440350-9903-000/​Objective-EC-Plan-Neofluar-40x-0.75|440350-9903-000]] ​ | +|  ​<fc #01b3f5>40x</​fc> ​ ​| ​ EC Plan-NeoFluar ​ |  Air  |  0.75  |  0.71  |  0.24  |  41 pixels ​ |  93.7 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440350-9903-000/​Objective-EC-Plan-Neofluar-40x-0.75|440350-9903-000]] ​ | 
-|  63x  |  Plan-Apochromat DIC  |  Oil  |  1.40  |  0.19  |  0.15  |  65 pixels ​ |  -  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420782-9900-000/​Objective-Plan-Apochromat-63x-1.4-Oil-DIC-M27|420782-9900-000]] ​ |+|  ​<fc #2c80c0>63x</​fc> ​ ​| ​ Plan-Apochromat DIC  |  Oil  |  1.40  |  0.19  |  0.15  |  65 pixels ​ |  -  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420782-9900-000/​Objective-Plan-Apochromat-63x-1.4-Oil-DIC-M27|420782-9900-000]] ​ |
 |  100x  |  α Plan-Apochromat DIC  |  Oil  |  1.46  |  0.11  |  0.09  |  105 pixels ​ |  -  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420792-9800-000/​Objective-alpha-Plan-Apochromat-100x-1.46-Oil-DIC-M27|420792-9800-000]] ​ | |  100x  |  α Plan-Apochromat DIC  |  Oil  |  1.46  |  0.11  |  0.09  |  105 pixels ​ |  -  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420792-9800-000/​Objective-alpha-Plan-Apochromat-100x-1.46-Oil-DIC-M27|420792-9800-000]] ​ |
  
Line 63: Line 63:
 **Upon request:** **Upon request:**
 ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  pixel size* (μm)  ^  10 μm =  ^  Reference ​ ^ ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  pixel size* (μm)  ^  10 μm =  ^  Reference ​ ^
-|  1.25x  |  EC Epiplan-Neofluar ​ |  Air  |  0.035  |  4  |  7.58  |  1.3 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​422310-9900-000/​Objective-EC-Epiplan-Neofluar-1.25x-0.03-M27|422310-9900-000]] ​ | +|  ​<fc #88888a>1.25x</​fc> ​ ​| ​ EC Epiplan-Neofluar ​ |  Air  |  0.035  |  4  |  7.58  |  1.3 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​422310-9900-000/​Objective-EC-Epiplan-Neofluar-1.25x-0.03-M27|422310-9900-000]] ​ | 
-|  5x  |  EC Plan-NeoFluar ​ |  Air  |  0.16  |  18.5  |  2.00  |  5 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420330-9901-000/​Objective-EC-Plan-Neofluar-5x-0.16-M27|420330-9901-000]] ​ | +|  ​<fc #df2b37>5x</​fc> ​ ​| ​ EC Plan-NeoFluar ​ |  Air  |  0.16  |  18.5  |  2.00  |  5 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420330-9901-000/​Objective-EC-Plan-Neofluar-5x-0.16-M27|420330-9901-000]] ​ | 
-|  40x  |  Plan-Neofluar Ph3  |  Oil  |  1.30  |  0.20  |  0.24  |  41 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440451-9903-000/​Objective-EC-Plan-Neofluar-40x-1.30-Oil-Ph3|440451-9903-000]] ​ | +|  ​<fc #01b3f5>40x</​fc> ​ ​| ​ Plan-Neofluar Ph3  |  Oil  |  1.30  |  0.20  |  0.24  |  41 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​440451-9903-000/​Objective-EC-Plan-Neofluar-40x-1.30-Oil-Ph3|440451-9903-000]] ​ | 
-|  63x  |  Plan-Apochromat Ph3  |  Oil  |  1.4  |  0.19  |  0.15  |  65 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420781-9910-000/​Objective-Plan-Apochromat-63-1.4-Oil-Ph3-M27|420781-9910-000]] ​ |+|  ​<fc #2c80c0>63x</​fc> ​ ​| ​ Plan-Apochromat Ph3  |  Oil  |  1.4  |  0.19  |  0.15  |  65 pixels ​ |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420781-9910-000/​Objective-Plan-Apochromat-63-1.4-Oil-Ph3-M27|420781-9910-000]] ​ |
  
 *** pixel sizes for binning = 1** *** pixel sizes for binning = 1**
Line 80: Line 80:
 Y = -8995 μm Y = -8995 μm
  
-===== System ​Turn On Procedures ​=====+===== Turn ON Procedure ​=====
   * Turn on the fluorescent light power source (if used)   * Turn on the fluorescent light power source (if used)
 {{halide_710.jpg?​80|}} {{halide_710.jpg?​80|}}
Line 93: Line 93:
   * Start the Metamorph software   * Start the Metamorph software
   * If the software doesn'​t detect the camera, power cycle the camera and try again   * If the software doesn'​t detect the camera, power cycle the camera and try again
-===== Microscope ​Turn Off procedures ​=====+  *  
 +===== Turn OFF procedure ​=====
  
 If there is another user for this microscope in the __next hour__: If there is another user for this microscope in the __next hour__:
Line 100: Line 101:
 Else: Else:
   * Clean up immersion objectives   * Clean up immersion objectives
-  * Log off the computer+  * Turn off the computer
   * Turn off the camera   * Turn off the camera
   * Turn off the Prior stage controller   * Turn off the Prior stage controller
zeiss_axiovert_200m.1765536805.txt.gz · Last modified: 2025/12/12 11:53 by bioimaging