User Tools

Site Tools


oeiras_zeiss_lumar_v12

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
oeiras_zeiss_lumar_v12 [2025/01/06 01:14]
bioimaging
oeiras_zeiss_lumar_v12 [2025/04/04 19:10] (current)
bioimaging [Pixel-size vs magnification]
Line 7: Line 7:
 <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​ <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​
 <​b>​Location</​b>:​ [Oeiras] Room 0B03<​br> ​ <​b>​Location</​b>:​ [Oeiras] Room 0B03<​br> ​
-<​b>​Manufacturer</​b>:​ <​br>​ +<​b>​Manufacturer</​b>: ​ZEISS <​br>​ 
-<​b>​Model</​b>:​ <​br>​ +<​b>​Model</​b>: ​Stereo Lumar V.12 <​br>​ 
-<​b>​Nickname</​b>:​ <​br>​ +<​b>​Nickname</​b>: ​Lumar <​br>​ 
-<​b>​Software</​b>:​ <​br>​ +<​b>​Software</​b>: ​Micromanager (Hamamatsu camera) / uEye (uEye camera) ​<​br>​ 
-<​b>​Year</​b>: ​ <br>+<​b>​Year</​b>: ​2011 <br>
 <​b>​SN</​b>: ​ <br> <​b>​SN</​b>: ​ <br>
 &nbsp <br> &nbsp <br>
Line 26: Line 26:
  
 ===== Microscope overview ===== ===== Microscope overview =====
 +
 +The Stereo Lumar V.12 is a motorized stereoscope with fluorescence,​ capable of imaging in colour and grayscale (camera dependent), with a zoom between 9.6x and 120x.
  
  
Line 31: Line 33:
  
 ^  Position ​ ^  Filterset ​ ^  Reference ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^ ^  Position ​ ^  Filterset ​ ^  Reference ​ ^  Excitation ​ ^  Dichroic ​ ^  Emission ​ ^
-   |    |  |  |  |  |+ ​1 ​ |  CFP  |  Filterset 47  |  436/​20 ​ |  455  |  480/​40 ​ | 
 +|  2  |  GFP  |  Filterset 13  |  470/​20 ​ |  495  |  505-530 ​ | 
 +|  3  |  TexasRed ​ |  Filterset 45  |  560/​40 ​ |  585  |  630/​75 ​ | 
 +|  4  |  YFP  |  Filterset 46  |  500/​20 ​ |  515  |  535/​30 ​ | 
 + 
 + 
 +===== Camera ===== 
 + 
 +Camera can be changed on request, and training for this can be provided if needed. 
 +^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^ 
 +|  [[https://​en.ids-imaging.com/​store/​ui-3370cp-rev-2.html|UI-3370CP Rev. 2 Colour]] ​ |  2048 × 2048  |  5.5 x 5.5  |  -  | 
 +|  [[https://​hep.hamamatsu.com/​eu/​en/​products/​C11440-42U30.html|ORCA-Flash4.0 LT+ Monochrome]] ​ |  2048 × 2048  |  6.5 x 6.5  |  82%  | 
 + 
 +===== Pixel-size vs magnification ===== 
 + 
 +^  Magnification ​ ^  Hamamatsu (µm/​pixel) ​ ^  IDS (µm/​pixel) ​  ^ 
 +|    ​9.6x  ​ ​10.90 ​ ​| ​ ​9.40 ​ | 
 +|   ​10.0x ​ ​| ​ ​10.49 ​ |  9.02  | 
 +|   ​20.0x ​ |  5.07  |  4.52  | 
 +|   ​30.0x ​ |  3.41  |  3.02  | 
 +|   ​40.0x ​ |  2.59  |  2.26  | 
 +|   ​50.0x ​ |  2.05  |  1.81  | 
 +|   ​60.0x ​ |  1.73  |  1.51  | 
 +|   ​70.0x ​ |  1.49  |  1.30  | 
 +|   ​80.0x ​ |  1.29  |  1.13  | 
 +|   ​90.0x ​ |  1.16  |  1.01  | 
 +|  100.0x ​ |  1.04  |  0.91  | 
 +|  110.0x ​ |  0.95  |  0.83  | 
 +|  120.0x ​ |  0.87  |  0.76  |
  
  
-==== Objectives ​==== +===== Microscope Turn On Procedure ===== 
- Magnification ​  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^ +  ​* Turn on the **powerbar below the stone table**. 
-|   ​| ​   |    |    |    |   +  * **If fluorescence is necessary**,​ turn on the **fluorescent lamp controller** on top of the desk. 
 +===== Microscope Turn Off Procedure ===== 
 +  * Turn on the **powerbar below the stone table**. 
 +  * **If fluorescence was used**, turn off the **fluorescent lamp controller** on top of the desk. 
 +    ​* This lamp **has to be on for 30 minutes**. If you use the system less than 30 minutes, **wait** until that time to turn the lamp off, or **warn the Bioimaging Team**.
  
-==== Camera ==== 
  
  
 ---- ----
 [[resources|Back to the Equipment page]] [[resources|Back to the Equipment page]]
oeiras_zeiss_lumar_v12.1736122455.txt.gz · Last modified: 2025/01/06 01:14 by bioimaging