bioimaging [System Specifications]
ivis_lumina_lt [2026/09/17 14:11] (current)
bioimaging [System overview]
Line 41: Line 41:
 {{:​warning.png?​25|}} Data files older than **3 months** will be automatically deleted on this system, please copy your data to the GIMM server using the desktop link. {{:​warning.png?​25|}} Data files older than **3 months** will be automatically deleted on this system, please copy your data to the GIMM server using the desktop link.
  
 +Additionally,​ IVIS Lumina LT is equipped with a [[https://​www.kentscientific.com/​products/​somnoflo/?​srsltid=AfmBOoo0Ou5Ldx5Qcyl9O-LhECWrG5Y_qDed5DOdSe9kc8OGR6y0p3su|SomnoFlo® anaesthesia system]] ​ for experiments with live animals using isoflurane. This system allows the use of smaller amounts of anaesthesia when carrying out live-imaging experiments,​ keeps mice stable during image acquisition,​ and reduces their recovery time, preventing the loss of animals. The setup is currently able to keep **up to three mice receiving vaporized isoflurane inside the IVIS Lumina LT**, while being imaged at the same time.
 + 
 +Users that are planning on using the anaesthesia system should contact the **Rodent Platform'​s designated veterinarian,​ Daniel Costa** (daniel.costa@gimm.pt),​ to schedule a training session.
 ===== System booking rules ===== ===== System booking rules =====
  
Line 49: Line 52:
   * **Camera sensor**: Back-thinned,​ back-illuminated CCD   * **Camera sensor**: Back-thinned,​ back-illuminated CCD
   * **CCD cooling temperature**:​ -90 °C   * **CCD cooling temperature**:​ -90 °C
-  * **Frame size**: 1024 x 1024 (default binning mode is 4x4 so acquired image is 256x256) ​- TO BE CONFIRMED!+  * **Frame size**: 1024 x 1024 (default binning mode is 4x4 so acquired image is 256x256) ​<= CONFIRM WITH STAFF!
   * **Pixel size**: 13 μm (52 in small binning mode)   * **Pixel size**: 13 μm (52 in small binning mode)
   * **Quantum Efficiency**:​ >85% Efficiency 500-700 nm, >55% Efficiency 400-500 nm, >35% Efficiency 700-900 nm   * **Quantum Efficiency**:​ >85% Efficiency 500-700 nm, >55% Efficiency 400-500 nm, >35% Efficiency 700-900 nm
Line 55: Line 58:
   * **Lens**: f/.95 – f/16, 50 mm   * **Lens**: f/.95 – f/16, 50 mm
   * **Sensitivity**:​ 100 photons/​s/​cm²/​sr   * **Sensitivity**:​ 100 photons/​s/​cm²/​sr
-  * **Read noise**: < 3 electrons for bin = 1, 2, 4; < 5 electrons for bin = 8, 16 <= CONFIRM ​THESE MODES+  * **Read noise**: < 3 electrons for bin = 1, 2, 4; < 5 electrons for bin = 8, 16 <= CONFIRM ​WITH STAFF!
   * **Dark Current**: < 120 electrons/​s/​cm2;​ or 2 x 10-4 electrons/​s/​pixel (typical)   * **Dark Current**: < 120 electrons/​s/​cm2;​ or 2 x 10-4 electrons/​s/​pixel (typical)
   * **Excitation fluorescence filters**: 10   * **Excitation fluorescence filters**: 10
   * **Emmission fluorescence filters**: 4   * **Emmission fluorescence filters**: 4
  
 +  * **Linear regime of camera is between 600 - 60 000 counts** (avoid operating above 60k!)
 + 