 This system was built to be versatile, and allow imaging of different sample sizes. It includes both high and low magnification objectives (which require different pinhole sizes, 25 or 40um). ​ This system was built to be versatile, and allow imaging of different sample sizes. It includes both high and low magnification objectives (which require different pinhole sizes, 25 or 40um). ​
 The scanhead allows 1x or 2x magnification (for extra magnification and to allow Nyquist sampling and deconvolution - see table below). ​ The scanhead allows 1x or 2x magnification (for extra magnification and to allow Nyquist sampling and deconvolution - see table below). ​
 +
 +Camera: [[https://​andor.oxinst.com/​assets/​uploads/​products/​andor/​documents/​andor-sona-scmos-specifications.pdf|Andor Sona-4BV11]]
 ===== System components ===== ===== System components =====
  
Line 60: Line 62:
 ==== Objectives ==== ==== Objectives ====
  
-^  Objectives ​ ^  NA  ^  Imm.  ^  WD (mm)  ^  Max FoV (1600x1350px) ​ ^  ideal CSU pinhole ​ ^  Theor. Resol. um  ^ Pixel size in μm (1x / 2x zoom)  ^  PFS Compatible ​ ^ +^  Objectives ​ ^  NA  ^  Imm.  ^  WD (mm)  ^  Max FoV (1600x1350px) ​ ^  ideal CSU pinhole ​ ^  Theor. Resol. um  ^  ​Measured pixel size in μm (1x zoom)  ^  PFS Compatible ​ ^ 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1825|10x]] ​ |  0.3  |  Dry  |  16  |  1.3mm  |  25um  |  1.07/​0.43 ​ |  1.020 / 0.510  ​|  Yes  | +|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1825|10x]] ​ |  0.3  |  Dry  |  16  |  1.3mm  |  25um  |  1.07/​0.43 ​ |  1.0561   |  Yes  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-179802|20x]] ​ |  0.8  |  Dry  |  0.8  |  0.68mm ​ |  40um  |  0.40/​0.16 ​ |  0.510 / 0.255  ​| ​ Yes  | +|  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-179802|20x]] ​ |  0.8  |  Dry  |  0.8  |  0.68mm ​ |  40um  |  0.40/​0.16 ​ |  0.5097  ​| ​ Yes  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-179806|40x]] ​ |  0.95  |  Dry  |  0.25-0.17 ​ |  0.34mm ​ |  40um  |  0.34/​0.13 ​ |  0.255 / 0.108* ​ ​| ​ Yes  | +|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-179806|40x]] ​ |  0.95  |  Dry  |  0.25-0.17 ​ |  0.34mm ​ |  40um  |  0.34/​0.13 ​ |  0.2580  ​| ​ Yes  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1855|60x]] ​ |  1.2  |  Water or Sil  |  0.31-0.28 ​ |  0.23mm ​ |  40um  |  0.27/​0.11 ​ |  0.170 / 0.085* ​ ​| ​ Yes  | +|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1855|60x]] ​ |  1.2  |  Water or Sil  |  0.31-0.28 ​ |  0.23mm ​ |  40um  |  0.27/​0.11 ​ |  0.1719  ​| ​ Yes  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1827|20x]]* ​ |  0.75  |  Multi  |  0.51-0.33 ​ |  0.68mm ​ |  25um  |  0.43/​0.17 ​ |  0.510 / 0.255  ​| ​ No  | +|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-1827|20x]]* ​ |  0.75  |  Multi  |  0.51-0.33 ​ |  0.68mm ​ |  25um  |  0.43/​0.17 ​ |  0.5097  ​| ​ No  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-5994|40x]]* ​ |  1.15  |  Water/​Sil ​ |  0.59-0.61 ​ |  0.34mm ​ |  40um  |  0.28/​0.11 ​ |  0.255 / 0.108* ​ ​| ​ Yes  | +|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-5994|40x]]* ​ |  1.15  |  Water/​Sil ​ |  0.59-0.61 ​ |  0.34mm ​ |  40um  |  0.28/​0.11 ​ |  0.2580  ​| ​ Yes  | 
-|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-179810|100x]]* ​ |  1.45  |  Oil  |  0.13  |  0.13mm ​ |  40um  |  0.22/​0.09 ​ |  0.102 / 0.050* ​ ​| ​ Yes  |+|  [[https://​www.microscope.healthcare.nikon.com/​pt_AMS/​products/​optics/​selector/​comparison/​-179810|100x]]* ​ |  1.45  |  Oil  |  0.13  |  0.13mm ​ |  40um  |  0.22/​0.09 ​ |  0.1038  ​| ​ Yes  |
  
 *Optional ​ *Optional ​
Line 73: Line 75:
 ===== Turn On Procedure ===== ===== Turn On Procedure =====
  
-1. Turn on the Computer+1. Turn on the Computer ​(if needed)
  
 2. Turn on power bar #1 and then power bar #2. These power bars are behind the microscope table, on a rail just above the floor. 2. Turn on power bar #1 and then power bar #2. These power bars are behind the microscope table, on a rail just above the floor.
dragonfly.1736867260.txt.gz · Last modified: 2025/01/14 16:07 by bioimaging