User Tools

Site Tools


andor_w1

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
andor_w1 [2025/01/15 15:00]
bioimaging [turning on]
andor_w1 [2025/01/15 15:06] (current)
bioimaging [CSU Dichroics]
Line 43: Line 43:
 ==== Objectives ==== ==== Objectives ====
  
-^  ​Magnification ​ ​^ ​ Model  ^  ​Immersion  ​^  NA  ^  WD (mm)  ^  PFS compatible ​ ​^ ​ DIC compatible ​ ​^ ​ PhC compatible ​ ​^ ​ Reference ​ ^+^  ​Mag.  ​^ ​ Model  ^  ​Imm. ^  NA  ^  WD (mm)  ^  PFS ​^ ​ DIC ​^ ​ PhC ​^ ​ Reference ​ ^
 |   ​4x ​ |  Nikon Plan DL   ​| ​ -  |  0.13  |  16.5  |  N  |  N/A  |  Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1834|MRH20045]] ​ |    |   ​4x ​ |  Nikon Plan DL   ​| ​ -  |  0.13  |  16.5  |  N  |  N/A  |  Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1834|MRH20045]] ​ |   
-|  10x  |  Nikon Plan DL  |  -  |  0.25  |  10.5  |  Yes  |  ​Not recommended ​ ​| ​ Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1701|MRL20102]] ​ |    +|  10x  |  Nikon Plan DL  |  -  |  0.25  |  10.5  |  Yes  |  ​No  ​| ​ Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1701|MRL20102]] ​ |    
-|  20x  |  Nikon Plan Fluor  |  -  |  0.5  |  2.1  |  Yes  |  ​Not recommended ​ ​| ​ Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1837|MRH10205]] ​ |   +|  20x  |  Nikon Plan Fluor  |  -  |  0.5  |  2.1  |  Yes  |  ​No  ​| ​ Ph1  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1837|MRH10205]] ​ |   
 |  20x  |  Nikon Plan Fluor  |  MImm  |  0.75  |  0.35  |  No  |  Yes (N1)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1827|MRH07241]] ​ |    |  20x  |  Nikon Plan Fluor  |  MImm  |  0.75  |  0.35  |  No  |  Yes (N1)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1827|MRH07241]] ​ |   
 |  40x  |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ |    |  40x  |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ |   
 |  63x  |  Nikon Plan APO VC  |  W  |  1.20  |  0.175  |  No  |  No  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1855|MRD07602]] ​ |    |  63x  |  Nikon Plan APO VC  |  W  |  1.20  |  0.175  |  No  |  No  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​products/​optics/​selector/​comparison/​-1855|MRD07602]] ​ |   
- 
-* Not fabricated anymore ; * * LWD - Long Working Distance  ​ 
  
 ==== CSU Dichroics ==== ==== CSU Dichroics ====
Line 58: Line 56:
 |  1  |  [[https://​www.idex-hs.com/​store/​product-detail/​ff685_di02_25x36/​fl-007099|IR mirror]] ​ |  685LP  | |  1  |  [[https://​www.idex-hs.com/​store/​product-detail/​ff685_di02_25x36/​fl-007099|IR mirror]] ​ |  685LP  |
 |  2  |  [[https://​www.idex-hs.com/​store/​product-detail/​di01_t405_488_568_647_13x15x0_5/​fl-006984|LD 405/​488/​561/​640]] ​ |  Quad mirror 80-85% transmission| |  2  |  [[https://​www.idex-hs.com/​store/​product-detail/​di01_t405_488_568_647_13x15x0_5/​fl-006984|LD 405/​488/​561/​640]] ​ |  Quad mirror 80-85% transmission|
-|  3  |  [[https://​www.chroma.com/​products/​parts/​zt405-488-561-640rpc|ZT 405/​488/​561/​640]] ​ |  Quad mirror >95% transmission |+|  3  |  [[https://​www.chroma.com/​products/​parts/​zt405-488-561-640rpc|ZT 405/​488/​561/​640]] ​ |  Quad mirror >95% transmission ​"​HE" ​|
  
  
Line 92: Line 90:
 ==== turning on ==== ==== turning on ====
  
-  * 1st: Turn on the cameras water cooling system (1 switches) and wait until it cools down to 15ºC. This should turn on both components of the water cooling system, if this does not happen, tell us as soon as possible! +  * 1st: Turn on the cameras water cooling system (1 switches) and wait until it cools down to 15ºC. This should turn on both components of the water cooling system, if this does not happen, tell us as soon as possible! **__Important__**:​ This should be done at least 10 to 15min before starting and you cannot start until it's at 15ºC.
-**__Important__**:​ This should be done at least 10 to 15min before starting and you cannot start until it's at 15ºC.+
   * Turn on the computer.   * Turn on the computer.
   * Log-In to the computer using your Agendo credentials (Username and Password).   * Log-In to the computer using your Agendo credentials (Username and Password).
andor_w1.1736949602.txt.gz · Last modified: 2025/01/15 15:00 by bioimaging