| 20x | Plan-Apochromat | Air | 0.80 | 0.55 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420650-9901-000/Objective-Plan-Apochromat-20x-0.8-M27|420650-9901-000]] | | | 20x | Plan-Apochromat | Air | 0.80 | 0.55 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420650-9901-000/Objective-Plan-Apochromat-20x-0.8-M27|420650-9901-000]] | | ||
| 40x | C-Apochromat Corr | Water | 1.20 | 0.28 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/421767-9970-000/Objective-C-Apochromat-40x-1.20-W-Corr-M27|421767-9970-000]] | | | 40x | C-Apochromat Corr | Water | 1.20 | 0.28 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/421767-9970-000/Objective-C-Apochromat-40x-1.20-W-Corr-M27|421767-9970-000]] | | ||
+ | | 40x | LD C-Apochromat Corr | Water | 1.10 | 0.62 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/421867-9970-000/Objective-LD-C-Apochromat-40x-1.1-W-Corr-M27|421867-9970-000]] | | ||
| 63x | Plan-Apochromat DIC | Oil | 1.40 | 0.19 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420782-9900-799/Objective-Plan-Apochromat-63x-1.4-Oil-DIC-M27|420782-9900-799]] | | | 63x | Plan-Apochromat DIC | Oil | 1.40 | 0.19 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420782-9900-799/Objective-Plan-Apochromat-63x-1.4-Oil-DIC-M27|420782-9900-799]] | | ||
Line 79: | Line 80: | ||
===== Microscope Turn On Procedure ===== | ===== Microscope Turn On Procedure ===== | ||
* Turn on the ''MAIN SWITCH'' | * Turn on the ''MAIN SWITCH'' | ||
- | * Turn on the ''SYSTEMS/PC'' switch | ||
{{zeiss_lsm_880_1.jpg?100}} | {{zeiss_lsm_880_1.jpg?100}} | ||
+ | * Turn on the incubation system (**A** on remote control) | ||
+ | * Turn on the ''SYSTEMS/PC'' switch | ||
* Turn on the computer | * Turn on the computer | ||
* Login in to Windows (Bioimaging User) | * Login in to Windows (Bioimaging User) | ||
Line 100: | Line 102: | ||
* Shutdown the computer | * Shutdown the computer | ||
* Turn off the two switches ''SYSTEMS/PC'' and ''COMPONENTS'' | * Turn off the two switches ''SYSTEMS/PC'' and ''COMPONENTS'' | ||
+ | * Turn off the incubation system (**A** on remote control) | ||
* **Wait 5 min. for Ar laser to cool down** | * **Wait 5 min. for Ar laser to cool down** | ||
* Turn off the ''MAIN SWITCH'' | * Turn off the ''MAIN SWITCH'' | ||
---- | ---- | ||
- | [[resources|Back to the Resources page]] | + | [[resources|Back to the Equipment page]] |