-===== What should I expect when sorting? ===== 
  
-==== Do I need to operate the equipment by myself? ==== 
- 
-We will train you to use the analyzers on your own. The sorters can only be operated by [about|Flow Cytometry Facility staff]]. You can book time with us and we will sort for you. But if you want to learn cell sorting, we will teach you. 
- 
-==== How are my cells collected? ==== 
- 
-{{  ::​how_many_cells.png?​400|}}You have many different options for collecting your sorted cells. 
- 
-  * Collect up to four different populations simultaneously into tubes of your choice, 1.5 mL, 5 mL tubes, or up to two population into 15 mL tubes. 
-  * Sort into any kind of plate. 6-well,​12-well,​ 24-well,​36-well,​ 96-well, 384-well, PCR plate, etc. 
- 
-Sort into the collection media of your choice. Collection tubes should be about 1/3 full of collection media. If you want to culture your sorted cells it’s a good idea to add Pen-Strep to the collection media. You can also sort directly into lysis buffer (for example, buffer that contains Trizol). 
- 
-We can adjust the temperature of the sorters to 4 ºC, 20 ºC, 37 ºC and 42 ºC, and we can also keep your cells on ice, if that is what you need. 
- 
-The BD FACSAria sorters faster with a smaller nozzle at higher pressure. With the 70 µm nozzle, you can sort up to 20,000 events/​second. This is a great choice for small cells that are happy in single-cell suspension, for example blood and bone marrow. Larger cells need a larger nozzle size. Your cells should be no more than one-third the size of the nozzle. Cells that are fragile or easily stressed (e.g. cultured cells) need a lower pressure sort. Our sorters have the following options: ​ 
-  * 70 µm nozzle / 70 psi 
-  * 85 µm nozzle / 45 psi 
-  * 100 µm nozzle / 20 psi 
- 
-We normally use saline (0.9% NaCl) in our sorters as sheath fluid. Each individual cell is sorted inside a tiny drop of this sheath fluid. If you are using the 70 µm nozzle (high pressure, faster sorting), you will get approximately 1 ml of sheath fluid with every 1 million sorted cells. If you are using the 100um nozzle (low pressure, slower sorting), you will get approximately 3 ml of sheath fluid with every 1 million sorted cells. 
- 
-[[newuser|Back to New User]] 