User Tools

Site Tools


quality

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
quality [2021/12/16 13:01]
flowcytometry
quality [2022/04/19 17:13] (current)
flowcytometry
Line 1: Line 1:
-====== ​Flow Cytometers ​Quality Control ======+====== ​Maintenance and Quality Control ====== 
 + 
 +{{  cs_t_beads.png?​700 ​ }}  
 + 
 +We perform daily **Cytometer Setup & Tracking (CST)** on our cytometers using BD CS&T research beads to track performance metrics such as detection efficiency, standard deviation of electronic noise, laser current, and optical background.\\ 
 +If there is a major alteration over time, we troubleshoot the problem to solve it as promptly as possible.  
 +\\ 
 +\\ 
 +{{  rainbow_beads.png?​1000 ​ }}  
 +\\ 
 +We also run **Rainbow 8 peak Beads (RB)** as a routine calibration of the flow cytometers to evaluate the linearity of the system.  
 +\\ 
 +These RB 8 peak beads are also specific for following the EuroFlow(TM) standardization guidelines.\\
  
-{{cs_t_beads.png |}} We perform daily **Cytometer Setup & Tracking (CST)** on our cytometers using BD CS&T research beads to track performance metrics such as detection efficiency, standard deviation of electronic noise, laser current, and optical background. We also run Rainbow Beads (RB) to track the performance of the BD LSRFortessa analyzers.\\ 
-You can check the CST results for each system on Diva (go to Cytometer -> Performance Tracking (LJ)).\\ 
 Please contact us if you want to check the CST or RB values for a particular date. Please contact us if you want to check the CST or RB values for a particular date.
 \\ \\
quality.1639656106.txt.gz · Last modified: 2021/12/16 13:01 by flowcytometry