This is an old revision of the document!
![]() |
Location: Oeiras Site, Room 0B01, Bartolomeu Dias Wing |
The BD LSRFortessa cell analyzer is a benchtop digital flow cytometer equipped with a fixed-alignment flow cell that is gel-coupled to the collection optics. It has three spatially separated lasers - 442 nm, 488 nm and 561 nm. Its primary function is to characterize complex populations of cells. Individual cells pass single file through a stream, which is interrogated by the three lasers with the emitted light being detected by photomultiplier (PMT) tubes. The system electronics calculates the time it takes a single cell to pass through each laser intercept and assigns the signal of each PMT to the appropriate cell. Besides fluorescence, the physical properties of the cell are also analyzed with side scatter (SSC), which reflects the internal complexity of a cell and forward scatter (FSC), which is roughly proportional to cell size. The sample flow rate can be set to low, medium and high values, which correspond approximately to 12, 35 and 60 μl/minute, respectively.
The LSRFortessa has a default configuration. Although, depending on the combination of fluorochromes in your samples, it allows the use of alternative configurations, using a set of optical filters and dichroic mirrors available.
The LSRFortessa is equipped with a High Throughput Sampler (HTS), which allows the automated acquisition of samples from 96 and 384 well plates.
Available 7 days a week, 24 hours a day for trained researchers.
Please check the User Guidelines for detailed information.