Line 10: Line 10:
 <​b>​Model</​b>:​ <a href="​https://​www.zeiss.com/​microscopy/​en/​products/​light-microscopes/​confocal-microscopes/​lsm-980-with-airyscan-2.html">​LSM 980 with Airyscan 2</​a><​br>​ <​b>​Model</​b>:​ <a href="​https://​www.zeiss.com/​microscopy/​en/​products/​light-microscopes/​confocal-microscopes/​lsm-980-with-airyscan-2.html">​LSM 980 with Airyscan 2</​a><​br>​
 <​b>​Nickname</​b>:​ "​980"<​br>​ <​b>​Nickname</​b>:​ "​980"<​br>​
-<​b>​Software</​b>:​ ZEN 3.(blue edition)<​br>​+<​b>​Software</​b>:​ ZEN 3.(blue edition)<​br>​
 <​b>​Year</​b>:​ 2021 <br> <​b>​Year</​b>:​ 2021 <br>
 <​b>​SN</​b>:​ 2841000239 <br> <​b>​SN</​b>:​ 2841000239 <br>
Line 29: Line 29:
 ===== Microscope overview ===== ===== Microscope overview =====
  
-{{  :​confocal_pinhole.png?​340}} The Zeiss LSM 980 with Airyscan 2 is a [[https://​www.youtube.com/​watch?​v=rTkjQ6Q0p5I|point-scanning confocal microscope]] able to generate high-resolution three-dimensional images of thick specimens with high sensitivity,​ high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments,​ equipped with a dark large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus.3 for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions, using the Tiles Advanced Setup of ZEN blue. Its [[http://​www.zeiss.com/​content/​dam/​Microscopy/​Products/​laser-scan/​LSM%20880/​zeiss-lsm880-airyscan-principle-672.jpg|Airyscan 2]] with 32 concentrically arranged GaAsP detection elements is able to collect all light from an Airy pattern simultaneously,​ allowing for increased resolution (120 nm laterally and 400 nm axially in SR mode, at 488 nm). It can also be used in [[https://​www.youtube.com/​watch?​v=tUiEa7vcggQ|Multiplex Mode]], which enables parallel excitation and detection in 4Y or 8Y modes. The result is a speed improvement by a factor of 4x/8x, while maintaining the sensitivity of Airyscan. Its scanning unit further includes two PMT detectors as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 488, 561 and 639 nm excitation wavelengths).+{{  :​confocal_pinhole.png?​340}} The Zeiss LSM 980 with Airyscan 2 is a [[https://​www.youtube.com/​watch?​v=rTkjQ6Q0p5I|point-scanning confocal microscope]] able to generate high-resolution three-dimensional images of thick specimens with high sensitivity,​ high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments,​ equipped with a dark large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus.3 for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions, using the Tiles Advanced Setup of ZEN blue. Its [[http://​www.zeiss.com/​content/​dam/​Microscopy/​Products/​laser-scan/​LSM%20880/​zeiss-lsm880-airyscan-principle-672.jpg|Airyscan 2]] with 32 concentrically arranged GaAsP detection elements is able to collect all light from an Airy pattern simultaneously,​ allowing for increased resolution (120 nm laterally and 350 nm axially in SR mode, at 488 nm). It can also be used in [[https://​www.youtube.com/​watch?​v=tUiEa7vcggQ|Multiplex Mode]], which enables parallel excitation and detection in 4Y or 8Y modes. The result is a speed improvement by a factor of 4x/8x, while maintaining the sensitivity of Airyscan. Its scanning unit further includes two PMT detectors as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 488, 561 and 639 nm excitation wavelengths). The [[https://​www.zeiss.com/​microscopy/​en/​about-us/​newsroom/​press-releases/​2021/​zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|LSM Plus deconvolution]] mode enhances signal-to-noise ratio by acting as a denoising step post image acquisition,​ allowing for high acquisition speed and low laser power. With [[https://​www.zeiss.com/​microscopy/​en/​about-us/​newsroom/​press-releases/​2021/​zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|Airyscan Joint Deconvolution (jDCV)]] the resolution can be improved to 90 nm (XY) and 270 nm (Z).
  
-  * [[https://​www.youtube.com/​watch?​v=zA4xqq-EL9g|ZEISS Webinar: ​Exploring the advancements in fast and gentle super-resolution imaging with LSM 980 Airyscan 2 Multiplex mode]]+  * [[https://​www.youtube.com/​watch?​v=zA4xqq-EL9g|ZEISS Webinar: LSM 980 Airyscan 2 Multiplex mode]]
  
 <​html>​ <​html>​
Line 62: Line 62:
 ==== LEDs (Ocular) ==== ==== LEDs (Ocular) ====
 ^  Line  ^  Wavelength ​ ^ ^  Line  ^  Wavelength ​ ^
-|  UV  |  ​405 nm  |+|  UV  |  ​385 nm  |
 |  B  |  475 nm  | |  B  |  475 nm  |
 |  G  |  555 nm  | |  G  |  555 nm  |