User Tools

Site Tools


zeiss_cell_observer

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
zeiss_cell_observer [2020/12/02 12:29]
bioimaging
zeiss_cell_observer [2021/01/17 04:44]
bioimaging [Microscope Turn Off procedures]
Line 48: Line 48:
 |  1  |  HE GFP  |  [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=489038-9901-000&​o=|FS38HE]] ​ |  450-490 nm  |  495 nm  |  500-550 nm  | |  1  |  HE GFP  |  [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=489038-9901-000&​o=|FS38HE]] ​ |  450-490 nm  |  495 nm  |  500-550 nm  |
 |  2  |  HE DsRed  |  [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=489043-9901-000&​o=|FS43HE]] ​ |  538-562 nm  |  570 nm  |  570-640 nm  | |  2  |  HE DsRed  |  [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=489043-9901-000&​o=|FS43HE]] ​ |  538-562 nm  |  570 nm  |  570-640 nm  |
-|  3  |  ​DAPI  ​| ​ [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=488049-9901-000&​o=|FS49]]  |  ​335-383 nm  |  ​395 nm  |  ​420-470 nm  | +|  3  |  ​Cy5  ​| ​ [[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=488050-9901-000&​o=|FS50]]  |  ​625-655 nm  |  ​660 nm  |  ​665-715 nm  | 
-|  4  |  ​Cy5  ​| ​ ​[[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=488050-9901-000&​o=|FS50]]  ​ ​625-655 nm  ​ 660 nm  |  665-715 nm  ​|+|  4  |  ​Analyzer DIC  |||||
 |  5  |  QUAD  |  [[https://​www.micro-shop.zeiss.com/​en/​us/​shop/​filterAssistant/​filtersets/​489090-9100-000|FS90HE]] |  370-400 nm  |  405 nm  |  410-440 nm  | |  5  |  QUAD  |  [[https://​www.micro-shop.zeiss.com/​en/​us/​shop/​filterAssistant/​filtersets/​489090-9100-000|FS90HE]] |  370-400 nm  |  405 nm  |  410-440 nm  |
 |  :::  |  :::  |  :::  |  450-488 nm  |  493 nm  |  499-529 nm  | |  :::  |  :::  |  :::  |  450-488 nm  |  493 nm  |  499-529 nm  |
 |  :::  |  :::  |  :::  |  540-570 nm  |  575 nm  |  580-605 nm  | |  :::  |  :::  |  :::  |  540-570 nm  |  575 nm  |  580-605 nm  |
 |  :::  |  :::  |  :::  |  615-648 nm  |  653 nm  |  659-759 nm  | |  :::  |  :::  |  :::  |  615-648 nm  |  653 nm  |  659-759 nm  |
-|  6  |  ​Empty  |||||+|  6  |  ​DAPI  ​| ​ ​[[https://​www.micro-shop.zeiss.com/?​s=23108437692794b&​l=en&​p=us&​f=f&​a=v&​b=f&​id=488049-9901-000&​o=|FS49]]  ​ ​335-383 nm  ​ 395 nm  |  420-470 nm  ​|
  
 ==== Objectives ==== ==== Objectives ====
Line 82: Line 82:
   * Turn on the temperature and CO2 controller (if you need them).   * Turn on the temperature and CO2 controller (if you need them).
 {{zeiss_axio_observer_temp.jpg?​80|}} {{zeiss_axio_observer_temp.jpg?​80|}}
-  * If you need CO2, open the valve +  * If you need CO2, open the main valve and the Cell Observer CO2 valve 
-{{zeiss_axio_observer_co2.jpg?80|}}+{{::​p3a10_co2_valves.png?200|}}
   * Turn on the motorized stage controller (**A** on remote control)   * Turn on the motorized stage controller (**A** on remote control)
 {{zeiss_axio_observer_a.jpg?​height=150|}} {{zeiss_axio_observer_a.jpg?​height=150|}}
Line 109: Line 109:
   * Turn off the motorized stage controller (**A** on remote control)   * Turn off the motorized stage controller (**A** on remote control)
   * Turn off the Temperature and CO2 controller (if used)   * Turn off the Temperature and CO2 controller (if used)
 +  * Close the Cell Observer CO2 valve (if used)
   * Turn off the HXP 120 fluorescent light source   * Turn off the HXP 120 fluorescent light source
  
 ---- ----
-[[resources|Back to the Resources ​page]]+[[resources|Back to the Equipment ​page]]