This is an old revision of the document!
![]() |
Location: [Oeiras] 0B06
|
A multi-photon, or 2-photon, microscope uses modules similar to those of a laser scanning confocal microscope: A laser source for sample excitation, a scanhead with galvo mirrors to scan the excitation beam, and photo-multiplier tubes to detect fluorescence signals. 2-photon imaging has advantages over confocal: the use of a near-infrared laser allows deeper penetration into tissues, the radiation is less energetic and potentially less phototoxic, and because the excitation effect is confined to the focal plane, de-scanning and pinholes are not necessary, so there are less light looses on the detection side.
This Prairie Ultima system is integrated with a custom modified fixed stage ready for intra-vital imaging, and 2photon is ideal for studies in thick tissue slices, in-vivo, or imaging of small organisms.
Click on the image on the right to see it in higher detail
Data files older than 3 months will be automatically deleted on this system, please copy your data to the institutional server using the desktop link.
Laser Unit | Wavelength |
---|---|
Coherent - Chameleon | 650 - 1050 nm |
Magnification | Model | Immersion | NA | WD (mm) | Reference |
---|---|---|---|---|---|
20x | Olympus (now Evident!) XLUM FL Plan | Water (dipping) | 1.00 | 2.00 | https://www.thorlabs.com/thorproduct.cfm?partnumber=N20X-PFH |
Filterset | Reference | Excitation | Dichroic | Emission | Fluorochromes |
---|---|---|---|---|---|
2p (pos1) | - | mirror | - | - | all |
Green (pos3) | - | blue | - | ~525 nm | GFP, FITC, Alexa488 |
Orange (pos4) | - | green | - | ~580 nm | TRITC, Alexa568 |
Blue (pos5) | - | violet | - | ~460 nm | DAPI |
Detector | Filter | Transmission | Fluorochromes |
---|---|---|---|
CH1 (PMT) | Red | 640/680 nm | Alexa647, Cy5 |
Ch2 (GaAsP) | Green | 570-620 nm | mCherry, Alexa568, Cy3 |
Ch3 (GaAsP) | Far Red | 500-550 nm | GFP, Alexa488, FITC |
Ch4 (PMT)* | Blue | 435-485 nm | DAPI*, CFP or SHG |
*Ch4 can be used either in fluorescence mode, or alternatively as a BF (“DODT”) channel or for SHG - ask staff
Upon request:
(requires changing filters and the dichroic beamsplitter)
UNDER CONSTRUCTION!!!
Transfer your data to the institutional server (data is deleted every month)
If there is another user for this microscope in the next hour:
Else:
standby