deltavision

This is an old revision of the document!


Location: [Oeiras] Room 0B02
Manufacturer: GE (Currently Leica)
Model: DeltaVision OMX-SIM
Nickname: "OMX"
Software: SoftWrx running on Linux
Year: 2017
SN: #
Funding: This equipment was purchased through ERC grant ref# ERC-2015-CoG-683258-CentrioleBirthDeath.
Operating conditions: Temperature & atmosphere control (37°C + Humidity + CO2)
 
 
 
 
Deltavision OMX-SIM Usage Statistics

Microscope overview

SIM microscopy is one of the imaging techniques that can overcome the limits of resolution of widefield fluorescence microscopes (~200nm), by using patterned (“structured”) illumination. The illumination patterns produce high order spatial information that can be extracted by computational algorithms in Fourier space. A typical SIM image is a reconstruction of several patterned images, where the illumination pattern changes in both angle or phase in order to extract almost twice the original resolution. In the DV OMX system, a total of 15 images are acquired for each image (5x phases x 3 angles). SIM works best with shallow samples, such as attached cells, and is less employed for tissue imaging. Optimal optical conditions are important to avoid reconstructions artifacts. When successful, SIM can provide valuable information about organelles, cytoskeleton and fine intra-cellular details. The speed of acquisition of the DV OMX system also allows live-cell imaging.

Suggestion for description in “Materials and Methods”: “Images were acquired on an Deltavision OMX system, equipped with 2 PCO Edge 5.5 sCMOS cameras, using the a 60x 1.42NA Oil immersion objetive, GFP + CY5 fluorescence filtersets and DIC optics. Images were deconvolved/reconstructed with Applied Precision's softWorx software.

System components

Lasers

Excitation Wavelength Power
Violet 405 nm 100mW
Blue 488 nm 100mW
Green 568 nm 100mW
Red 640 nm 100mW

Objectives

Magnification Immersion NA WD (mm) Image Pixel Size Reference
20x UPlan SAPO - 0.75 0.6 0.240
60x Plan APO N Oil 1.42 0.15 0.080

"Channels" & respective emission filters/cameras

Position Emission wavelength Dyes Camera
DAPI 435/31 DAPI, CFP, Hoescht… 1
GFP 528/48 GFP, Alx488… 1
RFP 609/37 RFP, Alx568… 2
Cy5 683/40 GCy5, Alx633… 2
DIC - Brighfield 1/2

Cameras (1 and 2)

2x PCO Edge 5.5 sCMOS 2560×2160 (note that each camera uses only 1024×1024 pixels to avoid field illumination problems and to allow simultaneous acquisition of 2 channels per camera)

Procedures

Turning ON

  • Turn on the main power switch at the top module
  • Turn on the laser chassis module (key switch)
  • If necessary turn on temp controller
  • Log on the OMX SR sworkstation (user: “worx” Password: “system serial number”
  • Start the AcquireSR software, and in the acquistion software click “Instruments|Status”
  • In the Instrument status dialog box, select “Restart Hardware” and wait.
  • Once hardware restart is finished, select “Reset All Cameras”, and wait until the camera SNs are reported in the log box.
  • Mount sample and start imaging.

Turning OFF

If you are NOT the last user for the day:

  • Transfer your data to the institutional server. Data will be deleted every month!
  • Close SoftWorx
  • Clean the objective and remove all your materials

Else:

  • Turn “OFF” the laser control and Hardware control.
  • Log off from the workstation

Back to the Equipment page