Location: [Oeiras] Room 0B05 |
The Andor Spinning W1 system is a spinning disk laser confocal microscope with a sensitive EMCCD camera suitable for fast in-vivo imaging. It's one of the fastest and most sensitive systems at the facility. The W1 also has installed microinjector and micromanipulation units and is ready for microfuildics experiments. Note that this system does not have widefield fluorescence, so samples can only be previewed in brightfield mode.
Type | Excitation line | Nominal power |
---|---|---|
Diode405 | 405nm | 100mW |
DPSS488 | 488nm | 100mW |
DPSS561 | 561nm | 50mW |
Diode640 | 640nm | 100mW |
Magnification | Model | Immersion | NA | WD (mm) | PFS compatible | DIC compatible | PhC compatible | Reference |
---|---|---|---|---|---|---|---|---|
4x | Nikon Plan DL | - | 0.13 | 16.5 | N | N/A | Ph1 | MRH20045 |
10x | Nikon Plan DL | - | 0.25 | 10.5 | Yes | Not recommended | Ph1 | MRL20102 |
20x | Nikon Plan Fluor | - | 0.5 | 2.1 | Yes | Not recommended | Ph1 | MRH10205 |
20x | Nikon Plan Fluor | MImm | 0.75 | 0.35 | No | Yes (N1) | No | MRH07241 |
40x | Plan Fluor | Oil | 1.3 | 0.24 | Yes | Yes (N2) | No | MRH01401 |
63x | Nikon Plan APO VC | W | 1.20 | 0.175 | No | No | No | MRD07602 |
* Not fabricated anymore ; * * LWD - Long Working Distance
Position | Name/reflection | Transmission |
---|---|---|
1 | IR mirror | 685LP |
2 | LD 405/488/561/640 | Quad mirror 80-85% transmission |
3 | ZT 405/488/561/640 | Quad mirror >95% transmission |
Position | Name/Description | Transmission | Dichroic |
---|---|---|---|
1 | Quad Semrock | 440/521/607/700-25 | 2 or 3[HE] |
2 | DAPI Semrock | 447/60 | 2 or 3[HE] |
3 | GFP Semrock | 525/30 | 2 or 3[HE] |
4 | RFP Semrock | 607/36 | 2 or 3[HE] |
5 | Cy5 Semrock | 685/40 | 2 or 3[HE] |
6 | Empty | - | - |
7 | Cy5LP Chroma HE | 655 LP | 2 or 3[HE] |
8 | RFP Chroma HE | 595/50 | 2 or 3[HE] |
9 | GFP Chroma HE | 525/50 | 2 or 3[HE] |
10 | DAPI Chroma HE | 450/50 | 2 or 3[HE] |
Model/type | Chip Size | Dexel Size | Quantum Efficiency |
---|---|---|---|
Andor iXon 888 | 1024 × 1024 | 13 um | >95% |
Andor Zyla 4.2 PLUS /sCMOS | 2048 × 2048 | 6.5 um | 82% |
IDS Monochrome UI-3370CP-M-GL | 2048 × 2048 | 5.5 um | 60% |
Piezo Z Stage: PI 737.2SL - range: 220µm - Resolution: 1 nm
Important: This should be done at least 10 to 15min before starting and you cannot start until it's at 15ºC.
* Remember to choose the proper calibration if doing montages/size measurements. * In iQ you need a different calibration for each objective and to check if whether you are using the 1,5x auxiliary Optovar magnification or not (selector in bottom right of the microscope front panel).
Else: