The Andor Spinning W1 system is a spinning disk laser confocal microscope with a sensitive EMCCD camera suitable for fast in-vivo imaging. It's one of the fastest and most sensitive systems at the facility. The W1 also has installed microinjector and micromanipulation units and is ready for microfuildics experiments. Note that this system does not have widefield fluorescence, so samples can only be previewed in brightfield mode.
Type | Excitation line | Nominal power |
---|---|---|
Diode405 | 405nm | 100mW |
DPSS488 | 488nm | 100mW |
DPSS561 | 561nm | 50mW |
Diode640 | 640nm | 100mW |
Magnification | Model | Immersion | NA | WD (mm) | PFS compatible | DIC compatible | PhC compatible | Reference |
---|---|---|---|---|---|---|---|---|
4x | Nikon Plan DL | - | 0.13 | 16.5 | N | N/A | Ph1 | MRH20045 |
10x | Nikon Plan DL | - | 0.25 | 10.5 | Yes | Not recommended | Ph1 | MRL20102 |
20x | Nikon Plan Fluor | - | 0.5 | 2.1 | Yes | Not recommended | Ph1 | MRH10205 |
20x | Nikon Plan Fluor | MImm | 0.75 | 0.35 | No | Yes (N1) | No | MRH07241 |
40x | Plan Fluor | Oil | 1.3 | 0.24 | Yes | Yes (N2) | No | MRH01401 |
63x | Nikon Plan APO VC | W | 1.20 | 0.175 | No | No | No | MRD07602 |
* Not fabricated anymore ; * * LWD - Long Working Distance
Position | Name/reflection | Transmission |
---|---|---|
1 | IR mirror | 685LP |
2 | LD 405/488/561/640 | Quad mirror 80-85% transmission |
3 | ZT 405/488/561/640 | Quad mirror >95% transmission |
Position | Name/Description | Transmission | Dichroic |
---|---|---|---|
1 | Quad Semrock | 440/521/607/700-25 | 2 or 3[HE] |
2 | DAPI Semrock | 447/60 | 2 or 3[HE] |
3 | GFP Semrock | 525/30 | 2 or 3[HE] |
4 | RFP Semrock | 607/36 | 2 or 3[HE] |
5 | Cy5 Semrock | 685/40 | 2 or 3[HE] |
6 | Empty | - | - |
7 | Cy5LP Chroma HE | 655 LP | 2 or 3[HE] |
8 | RFP Chroma HE | 595/50 | 2 or 3[HE] |
9 | GFP Chroma HE | 525/50 | 2 or 3[HE] |
10 | DAPI Chroma HE | 450/50 | 2 or 3[HE] |
Model/type | Chip Size | Dexel Size | Quantum Efficiency |
---|---|---|---|
Andor iXon 888 | 1024 × 1024 | 13 um | >95% |
Andor Zyla 4.2 PLUS /sCMOS | 2048 × 2048 | 6.5 um | 82% |
IDS Monochrome UI-3370CP-M-GL | 2048 × 2048 | 5.5 um | 60% |
Piezo Z Stage: PI 737.2SL - range: 220µm - Resolution: 1 nm
* 1st: Turn on the cameras water cooling system (1 switches) and wait until it cools down to 15ºC. This should turn on both components of the water cooling system, if this does not happen, tell us as soon as possible! Important: This should be done at least 10 to 15min before starting and you cannot start until it's at 15ºC. * Turn on the computer. * Log-In to the computer using your Agendo credentials (Username and Password). * Turn on the Main Cabinet key and switches. * Turn on the Power Bars. * Turn on the Spinning Disk scanhead by rotating the key. * Turn on the cameras. Only if the cooler is at 15ºC. * Press the button for the EMCCD and the switch for the sCMOS. * Start the iQ and input your iQ username and passowrd (it is different from the Agendo one!). If the software asks to initialize the stage, please do it BUT make sure the objective is “escaped” (click escape button on the right side of the microscope and when it's done click refocus to return to the previous focus). * Remember to choose the proper calibration if doing montages/size measurements. * In iQ you need a different calibration for each objective and to check if whether you are using the 1,5x auxiliary Optovar magnification or not (selector in bottom right of the microscope front panel).
Else: