Location: [Oeiras] Room 0B05
Manufacturer: Andor
Model: Revolution WD
Scanhead: Yokogowa W1
Nickname: "W1"
Software: Metamorph or IQ
Year: 2013
SN:
 
Data will be deleted after: 1 month  
 
Andor Spinning Disk W1 Quality Control
Andor Spinning Disk W1 Usage Statistics

Microscope overview

The Andor Spinning W1 system is a spinning disk laser confocal microscope with a sensitive EMCCD camera suitable for fast in-vivo imaging. It's one of the fastest and most sensitive systems at the facility. The W1 also has installed microinjector and micromanipulation units and is ready for microfuildics experiments. Note that this system does not have widefield fluorescence, so samples can only be previewed in brightfield mode.

System components

Lasers

Type Excitation line Nominal power
Diode405 405nm 100mW
DPSS488 488nm 100mW
DPSS561 561nm 50mW
Diode640 640nm 100mW

Objectives

Mag. Model Imm. NA WD (mm) PFS? DIC? PhC? Reference
4x Nikon Plan DL - 0.13 16.5 N N/A Ph1 MRH20045
10x Nikon Plan DL - 0.25 10.5 Yes No Ph1 MRL20102
20x Nikon Plan Fluor - 0.5 2.1 Yes No Ph1 MRH10205
20x Nikon Plan Fluor MImm 0.75 0.35 No Yes (N1) No MRH07241
40x Plan Fluor Oil 1.3 0.24 Yes Yes (N2) No MRH01401
63x Nikon Plan APO VC W 1.20 0.175 No No No MRD07602

CSU Dichroics

Position Name/reflection Transmission
1 IR mirror 685LP
2 LD 405/488/561/640 Quad mirror 80-85% transmission
3 ZT 405/488/561/640 Quad mirror >95% transmission “HE”

Filters (CSU Filterwheel)

Position Name/Description Transmission Dichroic
1 Quad Semrock 440/521/607/700-25 2 or 3[HE]
2 DAPI Semrock 447/60 2 or 3[HE]
3 GFP Semrock 525/30 2 or 3[HE]
4 RFP Semrock 607/36 2 or 3[HE]
5 Cy5 Semrock 685/40 2 or 3[HE]
6 Empty - -
7 Cy5LP Chroma HE 655 LP 2 or 3[HE]
8 RFP Chroma HE 595/50 2 or 3[HE]
9 GFP Chroma HE 525/50 2 or 3[HE]
10 DAPI Chroma HE 450/50 2 or 3[HE]

Camera(s)

Model/type Chip Size Dexel Size Quantum Efficiency
Andor iXon 888 1024 × 1024 13 um >95%
Andor Zyla 4.2 PLUS /sCMOS 2048 × 2048 6.5 um 82%
IDS Monochrome UI-3370CP-M-GL 2048 × 2048 5.5 um 60%

Piezo Z Stage: PI 737.2SL - range: 220µm - Resolution: 1 nm