zeiss_lsm_980 [2024/07/15 20:54] bioimaging [Microscope Turn On Procedure (after complete shutdown)] |
zeiss_lsm_980 [2025/02/18 10:21] (current) bioimaging [Microscope Turn On Procedure (after complete shutdown)] |
||
---|---|---|---|
Line 1: | Line 1: | ||
<html> | <html> | ||
- | <img src="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_name_new.png"> | + | <img src="/facility/bioimaging/lib/exe/fetch.php?media=lisbon_zeiss_lsm_980_with_airyscan_2_name_gimm_2024.png"> |
<table style="border:0px solid white;"> | <table style="border:0px solid white;"> | ||
<tr style="border:0px solid white;"> | <tr style="border:0px solid white;"> | ||
<td style="border:0px solid white;"> | <td style="border:0px solid white;"> | ||
- | <a href="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_wiki_2022.png"><img src="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_wiki_2022.png" width=300></a> | + | <a href="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_wiki_2024.png"><img src="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_wiki_2024.png" width=300></a> |
<td style="border:0px solid white;"> <p style="line-height:1.8"> | <td style="border:0px solid white;"> <p style="line-height:1.8"> | ||
- | <b>Location</b>: <a href="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_map_new_p2a24.png">Room P2-A-24</a> (<img src="/facility/bioimaging/lib/exe/fetch.php?media=phone_neg.png" width=15> 47219) <br> | + | <b>Location</b>: <a href="/facility/bioimaging/lib/exe/fetch.php?media=zeiss_lsm_980_map_new_p2a24.png">[Lisbon] Room P2-A-24</a> (<img src="/facility/bioimaging/lib/exe/fetch.php?media=phone_neg.png" width=15> 47219) <br> |
<b>Manufacturer</b>: <a href="http://www.zeiss.de/micro" target="_blank">ZEISS Microscopy</a><br> | <b>Manufacturer</b>: <a href="http://www.zeiss.de/micro" target="_blank">ZEISS Microscopy</a><br> | ||
<b>Model</b>: <a href="https://www.zeiss.com/microscopy/en/products/light-microscopes/confocal-microscopes/lsm-980-with-airyscan-2.html">LSM 980 with Airyscan 2</a><br> | <b>Model</b>: <a href="https://www.zeiss.com/microscopy/en/products/light-microscopes/confocal-microscopes/lsm-980-with-airyscan-2.html">LSM 980 with Airyscan 2</a><br> | ||
<b>Nickname</b>: "980"<br> | <b>Nickname</b>: "980"<br> | ||
- | <b>Software</b>: ZEN 3.8 (blue edition)<br> | + | <b>Software</b>: ZEN 3.8<br> |
<b>Year</b>: 2021 <br> | <b>Year</b>: 2021 <br> | ||
<b>SN</b>: 2841000239 <br> | <b>SN</b>: 2841000239 <br> | ||
  <br> |   <br> | ||
Data will be deleted after: <b>1 month</b> | Data will be deleted after: <b>1 month</b> | ||
- |   <br> | ||
  <br> |   <br> | ||
  <br> |   <br> | ||
Line 29: | Line 28: | ||
===== Microscope overview ===== | ===== Microscope overview ===== | ||
- | {{ :confocal_pinhole.png?340}} The Zeiss LSM 980 with Airyscan 2 is a [[https://www.youtube.com/watch?v=rTkjQ6Q0p5I|point scanning confocal microscope]] able to generate high-resolution three-dimensional images of thick specimens with high sensitivity, high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments, equipped with a dark large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus.3 for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions, using the Tiles Advanced Setup of ZEN blue. Its [[http://www.zeiss.com/content/dam/Microscopy/Products/laser-scan/LSM%20880/zeiss-lsm880-airyscan-principle-672.jpg|Airyscan 2]] with 32 concentrically arranged GaAsP detection elements is able to collect all light from an Airy pattern simultaneously, allowing for increased resolution (120 nm laterally and 350 nm axially in SR mode, at 488 nm). It can also be used in [[https://www.youtube.com/watch?v=tUiEa7vcggQ|Multiplex Mode]], which enables parallel excitation and detection in 4Y or 8Y modes. The result is a speed improvement by a factor of 4x/8x, while maintaining the sensitivity of Airyscan. Its scanning unit further includes two PMT detectors as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 488, 561 and 639 nm excitation wavelengths). The [[https://www.zeiss.com/microscopy/en/about-us/newsroom/press-releases/2021/zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|LSM Plus deconvolution]] mode enhances signal-to-noise ratio by acting as a denoising step post image acquisition, allowing for high acquisition speed and low laser power. With [[https://www.zeiss.com/microscopy/en/about-us/newsroom/press-releases/2021/zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|Airyscan Joint Deconvolution (jDCV)]] the resolution can be improved to 90 nm (XY) and 270 nm (Z). | + | {{ :confocal_pinhole.png?340}} The Zeiss LSM 980 with Airyscan 2 is a [[https://www.youtube.com/watch?v=rTkjQ6Q0p5I|point scanning confocal microscope]] able to generate high-resolution three-dimensional images of thick specimens with high sensitivity, high speed and low photodamage. It is an inverted microscope especially suitable for live cell imaging and photobleaching experiments, equipped with a dark large cage incubator and a stage top incubator for temperature control and CO2 supply and with Definite Focus.3 for hardware focus control during long time-lapse acquisitions. The stage is motorized and furthermore equipped with a piezo for Z displacement so fast 4D imaging is possible in multiple stage positions, using the Tiles Advanced Setup of ZEN blue. Its Airyscan 2 with 32 concentrically arranged GaAsP detection elements is able to collect all light from an Airy pattern simultaneously, allowing for increased resolution (120 nm laterally and 350 nm axially in SR mode, at 488 nm). It can also be used in [[https://www.youtube.com/watch?v=tUiEa7vcggQ|Multiplex Mode]], which enables parallel excitation and detection in 4Y or 8Y modes. The result is a speed improvement by a factor of 4x/8x, while maintaining the sensitivity of Airyscan. Its scanning unit further includes two PMT detectors as well as a GaAsP detector for increased sensitivity (45% QE compared to ~25% QE for conventional PMT). It is equipped with lasers from violet to far red (405, 488, 561 and 639 nm excitation wavelengths). The [[https://www.zeiss.com/microscopy/en/about-us/newsroom/press-releases/2021/zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|LSM Plus deconvolution]] mode enhances signal-to-noise ratio by acting as a denoising step post image acquisition, allowing for high acquisition speed and low laser power. With [[https://www.zeiss.com/microscopy/en/about-us/newsroom/press-releases/2021/zeiss-lsm-plus-and-zeiss-airyscan-joint-deconvolution.html|Airyscan Joint Deconvolution (jDCV)]] the resolution can be improved to 90 nm (XY) and 270 nm (Z). |
* **Microscope**: [[https://www.zeiss.com/microscopy/en/products/light-microscopes/widefield-microscopes/axio-observer-for-life-science-research.html|Zeiss Axio Observer]] | * **Microscope**: [[https://www.zeiss.com/microscopy/en/products/light-microscopes/widefield-microscopes/axio-observer-for-life-science-research.html|Zeiss Axio Observer]] | ||
Line 45: | Line 44: | ||
* [[https://www.youtube.com/watch?v=zA4xqq-EL9g|ZEISS Webinar: LSM 980 Airyscan 2 Multiplex mode]] | * [[https://www.youtube.com/watch?v=zA4xqq-EL9g|ZEISS Webinar: LSM 980 Airyscan 2 Multiplex mode]] | ||
- | {{:warning.png?25|}} Data files older than **1 month** will be automatically deleted on this system, please copy your data to the iMM server using the desktop link. | + | {{:warning.png?25|}} Data files older than **1 month** will be automatically deleted on this system, please copy your data to the GIMM server using the desktop link. |
===== Booking Rules ===== | ===== Booking Rules ===== | ||
Line 85: | Line 84: | ||
===== How to start your session ===== | ===== How to start your session ===== | ||
- | The system should be always on, you only need to turn on the computer, log into Windows with your Agendo credentials and start **ZEN**. | + | The system should be always on, you only need to: |
+ | * Turn on the computer | ||
+ | * Log into Windows with your Agendo credentials | ||
+ | * Start **ZEN** | ||
===== How to end your session ===== | ===== How to end your session ===== | ||
Line 107: | Line 109: | ||
* Turn on the ''COMPONENTS'' switch | * Turn on the ''COMPONENTS'' switch | ||
{{::zeiss_lsm_980_components.png?100|}} | {{::zeiss_lsm_980_components.png?100|}} | ||
+ | * Turn on the Laser key | ||
* Turn on the computer | * Turn on the computer | ||
* Log into Windows with your Agendo credentials | * Log into Windows with your Agendo credentials | ||
Line 118: | Line 121: | ||
* Exit the ZEN software | * Exit the ZEN software | ||
* Shutdown the computer | * Shutdown the computer | ||
- | * Turn off the ''COMPONENTS'' switch (do not turn off the Laser key!) | + | * Turn off the Laser key |
+ | * Turn off the ''COMPONENTS'' switch | ||
* Turn off the ''MAIN SWITCH'' | * Turn off the ''MAIN SWITCH'' | ||
---- | ---- | ||
[[resources|Back to the Equipment page]] | [[resources|Back to the Equipment page]] |