User Tools

Site Tools


nikon_htm-hcs

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
nikon_htm-hcs [2025/12/12 10:14]
bioimaging
nikon_htm-hcs [2026/09/03 12:15] (current)
bioimaging
Line 4: Line 4:
 <tr style="​border:​0px solid white;">​ <tr style="​border:​0px solid white;">​
 <td style="​border:​0px solid white;">​ <td style="​border:​0px solid white;">​
-<a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=nikon_htm-hcs.jpg"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=nikon_htm-hcs.jpg"​ width=300></​a>​+<a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=oeiras_nikon_htm-hcs.jpg"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=oeiras_nikon_htm-hcs.jpg"​ width=300></​a>​
 <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​ <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​
-<​b>​Location</​b>:​ [Oeiras] Room 0B05<​br> ​+<​b>​Location</​b>: ​<a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=bartolomeu_dias_wing_2025_-_nikon_htm-hcs.png">​[Oeiras] Room 0B05</a><​br> ​
 <​b>​Manufacturer</​b>:​ <a href="​https://​www.microscope.healthcare.nikon.com/​en_EU/"​ target="​_blank">​Nikon</​a>​ <br> <​b>​Manufacturer</​b>:​ <a href="​https://​www.microscope.healthcare.nikon.com/​en_EU/"​ target="​_blank">​Nikon</​a>​ <br>
 <​b>​Model</​b>:​ <a href="​https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​inverted-microscopes/​eclipse-ti-series"​ target="​_blank">​Nikon Eclipse Ti</​a>​ <br> <​b>​Model</​b>:​ <a href="​https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​inverted-microscopes/​eclipse-ti-series"​ target="​_blank">​Nikon Eclipse Ti</​a>​ <br>
Line 30: Line 30:
 The Nikon HTM-HCS is a fully automated microscope capable of high-throughput imaging and screening of samples (eg, Ibidi chamber slides or well-plates);​ with the help of fully programmable and customizable acquisition protocols. ​ The Nikon HTM-HCS is a fully automated microscope capable of high-throughput imaging and screening of samples (eg, Ibidi chamber slides or well-plates);​ with the help of fully programmable and customizable acquisition protocols. ​
  
-The system is equipped with a CoolLED fluorescence illuminator with 4 excitation bands, and an Andor Zyla 4.sCMOS camera with 4.0 MPixels with a large field of view capable of maximum of 100 FPS at full frame (2048x2048)+The system is equipped with a CoolLED fluorescence illuminator with 4 excitation bands, and an [[https://​www.biovis.com/​images/​cameras/​SCAS0122E_C11440-42U30.pdf|ORCA-Flash4.0 LT+ sCMOS camera]] with 4.0 MPixels ​(2048x2048) ​with a large field of view, low-light sensitivity,​ and large dynamic range
  
 The optics available include fluorescence filters and brightfield,​ including DIC (for several magnifications) and phase-contrast (40x LWD only). The optics available include fluorescence filters and brightfield,​ including DIC (for several magnifications) and phase-contrast (40x LWD only).
Line 89: Line 89:
 |  <fc #​01b3f5>​40x</​fc> ​ |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ |  0.162  |  0.108  |    |  <fc #​01b3f5>​40x</​fc> ​ |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ |  0.162  |  0.108  |   
 |  <fc #​01b3f5>​40x</​fc> ​ |  Plan Apo  |  Air  |  0.95  |  0.17-0.25 ​ |  Yes  |  Yes (N2)  |  No  |  MRD00400* ​ |  0.161  |  0.107  |    |  <fc #​01b3f5>​40x</​fc> ​ |  Plan Apo  |  Air  |  0.95  |  0.17-0.25 ​ |  Yes  |  Yes (N2)  |  No  |  MRD00400* ​ |  0.161  |  0.107  |   
-|  <fc #​01b3f5>​40x</​fc> ​ |  ​LWD* *  |  Air  |  0.55  ​| ​ 2. ​|  ​No  ​| ​ No  |  Yes  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1683|MRP46402]]  |  0.161  |  0.108  |   +|  <fc #​01b3f5>​40x</​fc> ​ |  ​ELWD* *  |  Air  |  0.60  ​| ​ 2.80-3.60 ​ ​|  ​Yes  ​| ​ No  |  Yes  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1814|MRH48430]]  |  0.161  |  0.108  |   
 |  100x  |  Plan Apo Lambda ​ |  Oil  |  1.45  |  0.13  |  Yes  |  Yes (N2)  |  No  |  MRD01905* ​ |  0.065  |  0.043  |    |  100x  |  Plan Apo Lambda ​ |  Oil  |  1.45  |  0.13  |  Yes  |  Yes (N2)  |  No  |  MRD01905* ​ |  0.065  |  0.043  |   
  
Line 99: Line 99:
  
 ^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^ ^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^
-|  [[https://andor.oxinst.cn/assets/uploads/products/​andor/​documents/​Zyla-4.2-Specifications-Life-Science.pdf|Andor Zyla 4.PLUS]] ​ |  2048 × 2048  |  6.5 x 6.5  |  82%  |+|  [[https://www.biovis.com/images/cameras/SCAS0122E_C11440-42U30.pdf|ORCA-Flash4.0 LT PLUS]] ​ |  2048 × 2048  |  6.5 x 6.5  |  82%  |
  
 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
  
   * Turn on the microscope using the **switch** on the **bottom of the microscope'​s optical table**.   * Turn on the microscope using the **switch** on the **bottom of the microscope'​s optical table**.
-{{nikon-htm_turnon_3.png?​200|}}+{{nikon-htm_turnon_3.png?​150|}} 
 +  * Turn on the computer  
 +  * Log In into Windows with your Agendo credentials 
 +  * Start the **NIS** software
  
 ===== Microscope Turn Off Procedure ===== ===== Microscope Turn Off Procedure =====
 +If there is another user for this microscope in the next hour:
 +
 +  * Exit the **NIS** software
 +  * Log off the computer
 +  * Clean up immersion objectives  ​
 +
 +Else:
 +
 +  * Exit the **NIS** software
 +  * Clean up immersion objectives
 +  * Turn off the computer
   * Turn off the microscope using the **switch** on the **bottom of the microscope'​s optical table**.   * Turn off the microscope using the **switch** on the **bottom of the microscope'​s optical table**.
-{{nikon-htm_turnon_3.png?​200|}}+{{nikon-htm_turnon_3.png?​150|}}
  
 ---- ----
 [[resources|Back to the Equipment page]] [[resources|Back to the Equipment page]]
nikon_htm-hcs.1765530845.txt.gz · Last modified: 2025/12/12 10:14 by bioimaging