bioimaging [Microscope overview]
Line 28: Line 28:
 ===== Microscope overview ===== ===== Microscope overview =====
  
-The Nikon-HTM/HCS microscope ​is a fully automated microscope capable of high-throughput imaging and screening of samples (eg, Ibidi chamber slides or well-plates);​ with the help of fully programmable and customizable acquisition protocols. ​+The Nikon HTM-HCS is a fully automated microscope capable of high-throughput imaging and screening of samples (eg, Ibidi chamber slides or well-plates);​ with the help of fully programmable and customizable acquisition protocols. ​
  
 The system is equipped with a CoolLED fluorescence illuminator with 4 excitation bands, and an Andor Zyla 4.2 sCMOS camera with 4.0 MPixels with a large field of view capable of maximum of 100 FPS at full frame (2048x2048). ​ The system is equipped with a CoolLED fluorescence illuminator with 4 excitation bands, and an Andor Zyla 4.2 sCMOS camera with 4.0 MPixels with a large field of view capable of maximum of 100 FPS at full frame (2048x2048). ​
Line 62: Line 62:
 |  Cy5-FW ​ |  [[https://​www.chroma.com/​products/​parts/​s685-40m|s685-40m]] ​ |  685/​40 ​ | |  Cy5-FW ​ |  [[https://​www.chroma.com/​products/​parts/​s685-40m|s685-40m]] ​ |  685/​40 ​ |
 ===== Objectives ===== ===== Objectives =====
-^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  PFS compatible ​ ​^ ​ DIC compatible ​ ​^ ​ PhC compatible ​ ​^ ​ Reference ​ ^ +^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  PFS  ^  DIC  ^  PhC  ^  Reference ​ ​^ ​ Pixel size (um) (1x zoom)  ^  Pixel size (um) (1.5x zoom)  ^ 
-|  10x  |  Plan Apo  |  Air  |  0.45  |  4  |  No  |  Yes (N1)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-179798|MRD70170]] ​ |    +|  10x  |  Plan Apo  |  Air  |  0.45  |  4  |  No  |  Yes (N1)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-179798|MRD70170]] ​ ​| ​ 0.643  |  0.429  ​| ​   
-|  20x  |  Plan Apo Lambda ​ |  Air  |  0.75  |  1  |  Yes  |  Yes (N1)  |  No  |  MRD00205* ​ |    +|  20x  |  Plan Apo Lambda ​ |  Air  |  0.75  |  1  |  Yes  |  Yes (N1)  |  No  |  MRD00205* ​ ​| ​ 0.323  |  0.216  ​| ​   
-|  100x  |  Plan Apo Lambda ​ |  Oil  |  1.45  |  0.13  |  Yes  |  Yes (N2)  |  No  |  MRD01905* ​ |    +|  100x  |  Plan Apo Lambda ​ |  Oil  |  1.45  |  0.13  |  Yes  |  Yes (N2)  |  No  |  MRD01905* ​ ​| ​ 0.065  |  0.043  ​| ​   
-|  40x  |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ |    +|  40x  |  Plan Fluor  |  Oil  |  1.3  |  0.24  |  Yes  |  Yes (N2)  |  No  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1829|MRH01401]] ​ ​| ​ 0.162  |  0.108  ​| ​   
-|  40x  |  Plan Apo  |  Air  |  0.95  |  0.17-0.25 ​ |  Yes  |  Yes (N2)  |  No  |  MRD00400* ​ |    +|  40x  |  Plan Apo  |  Air  |  0.95  |  0.17-0.25 ​ |  Yes  |  Yes (N2)  |  No  |  MRD00400* ​ ​| ​ 0.161  |  0.107  ​| ​   
-|  40x  |  LWD* *  |  Air  |  0.55  |  2.1  |  No  |  No  |  Yes  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1683|MRP46402]] ​ |   +|  40x  |  LWD* *  |  Air  |  0.55  |  2.1  |  No  |  No  |  Yes  |  [[https://​www.microscope.healthcare.nikon.com/​en_EU/​products/​optics/​selector/​comparison/​-1683|MRP46402]] ​ ​| ​ 0.161  |  0.108  ​| ​  
  
 * Not fabricated anymore ; * * LWD - Long Working Distance  ​ * Not fabricated anymore ; * * LWD - Long Working Distance  ​
 +
 +The Bioimaging Platform **highly advises against using the 40x LWD objective with 1.5x zoom**