User Tools

Site Tools


deltavision

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
deltavision [2025/01/11 19:13]
bioimaging
deltavision [2025/12/12 11:03] (current)
bioimaging
Line 6: Line 6:
 <a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=deltavision.jpg"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=deltavision.jpg"​ width=500></​a>​ <a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=deltavision.jpg"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=deltavision.jpg"​ width=500></​a>​
 <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​ <td style="​border:​0px solid white;">​ <p style="​line-height:​1.8">​
-<​b>​Location</​b>:​ [Oeiras] Room 0B02<​br> ​+<​b>​Location</​b>: ​<a href="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=bartolomeu_dias_wing_2025_-_deltavision.png">​[Oeiras] Room 0B02</a><​br> ​
 <​b>​Manufacturer</​b>:​ GE (Currently Leica) <br> <​b>​Manufacturer</​b>:​ GE (Currently Leica) <br>
 <​b>​Model</​b>: ​ DeltaVision OMX-SIM <br> <​b>​Model</​b>: ​ DeltaVision OMX-SIM <br>
Line 34: Line 34:
 ==== Lasers ==== ==== Lasers ====
 ^  Excitation ​ ^  Wavelength ​ ^  Power  ^ ^  Excitation ​ ^  Wavelength ​ ^  Power  ^
-|  Violet ​ |  405 nm  |  100mW  | +|  Violet ​ |  ​<fc #a600ff>405 nm</​fc> ​ ​| ​ 100mW  | 
-|  Blue  |  488 nm  |  100mW  | +|  Blue  |  ​<fc #00f7ff>488 nm</​fc> ​ ​| ​ 100mW  | 
-|  Green  |  568 nm  |  100mW  | +|  ​Yellow-Green  |  ​<fc #dbff00>568 nm</​fc> ​ ​| ​ 100mW  | 
-|  Red  |  640 nm  |  100mW  |+|  Red  |  ​<fc #ff2100>640 nm</​fc> ​ ​| ​ 100mW  |
  
 ==== Objectives ==== ==== Objectives ====
-^  Magnification ​ ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Image Pixel Size  ​^ ​ Reference ​ ^ +^  Magnification ​ ​^ ​ Model  ​^ ​ Immersion ​ ^  NA  ^  WD (mm)  ^  Image Pixel Size  ^ 
-|  20x  UPlan SAPO  |  -  |  0.75  |  0.6  |  0.240  |  +|  ​<fc #0aa05c>20x</​fc> ​ |  UPlan SAPO  |  -  |  0.75  |  0.6  |  0.240  |  
-|  60x  Plan APO N  |  Oil  |  1.42  |  0.15  |   ​0.080* ​ |+|  ​<fc #2c80c0>60x</​fc> ​ |  Plan APO N  |  Oil  |  1.42  |  0.15  |   ​0.080* ​ |
  
 * after SIM reconstruction this system is capable of achieving resolutions of 100-150nm! * after SIM reconstruction this system is capable of achieving resolutions of 100-150nm!
  
-==== Emission filters/​cameras ​==== +==== Emission filters ==== 
-^  Channel ​ ^  ​Emission wavelength ​ ^  Dyes  ​^ ​ Camera ​ ^   + 
-|  DAPI  |  ​435/31  |  DAPI, CFP, Hoescht... ​ ​| ​ 1  |  +^  Channel ​ ^  ​Wavelength ​ ​^ ​ Camera ​ ^   
-|  GFP  |  ​528/48  |  GFP, Alx488... ​ ​| ​ 1  |     +|  DAPI  |  ​<fc #​2300ff>​420-450 nm</fc> ​ ​| ​ 1  |  
-|  RFP  |  ​609/37  |  RFP, Alx568... ​ ​| ​ 2  | +|  GFP  |  ​<fc #​56ff00>​504-552 nm</fc> ​ ​| ​ 1  |     
-|  Cy5  |  ​683/40  |  GCy5, Alx633... ​ ​| ​ 2  | +|  RFP  |  ​<fc #​ff9f00>​590-627 nm</fc> ​ ​| ​ 2  | 
-|  DIC*  |  -  ​| ​ Brighfield ​ ​| ​ 1/2  |+|  Cy5  |  ​<fc #​ff0000>​663-703 nm</fc> ​ ​| ​ 2  | 
 +|  ​Brightfield / DIC*  |  -  |  1/2  |
  
 * though possible, DIC imaging is not recommended as it may introduce artifacts in the fluorescence reconstructions * though possible, DIC imaging is not recommended as it may introduce artifacts in the fluorescence reconstructions
Line 60: Line 61:
 2x PCO Edge 5.5 sCMOS 2560x2160 (note that each camera uses only 1024x1024 pixels to avoid field illumination problems and to allow simultaneous acquisition of 2 channels per camera) 2x PCO Edge 5.5 sCMOS 2560x2160 (note that each camera uses only 1024x1024 pixels to avoid field illumination problems and to allow simultaneous acquisition of 2 channels per camera)
  
-===== Operating Procedures ===== +===== Turning ON =====
- +
-For more detailed information consult the full SOP/Guide here: +
-(under construction) +
- +
-==== Turning ON ====+
  
   * Turn on the main power switch at the top module   * Turn on the main power switch at the top module
Line 76: Line 72:
   * Mount sample and start imaging.   * Mount sample and start imaging.
  
-==== Turning OFF ====+===== Turning OFF =====
  
 If you are NOT the last user for the day: If you are NOT the last user for the day:
deltavision.1736619215.txt.gz · Last modified: 2025/01/11 19:13 by bioimaging