blaze

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
blaze [2026/02/17 17:03]
bioimaging
blaze [2026/05/26 11:47] (current)
bioimaging [Microscope Turn On Procedure]
Line 36: Line 36:
 ===== Laser Lines ===== ===== Laser Lines =====
  
-^  Wavelength ​(nm)  ​^ ​ Fluorophores ​ ^  Penetration capability ​ ^ +^  Wavelength ​ ^  Fluorophores ​ ^  Penetration capability ​ ^ 
-|  488  |  E.g: GFP, TRITC, AL488  |  Less penetration ​ | +|  ​<fc #00f7ff>488 nm</​fc> ​ ​| ​ E.g: GFP, TRITC, AL488  |  Less penetration ​ | 
-|  561  |  E.g: mCherry, RFP, AL594  |  Good penetration ​ | +|  ​<fc #c6ff00>561 nm</​fc> ​ ​| ​ E.g: mCherry, RFP, AL594  |  Good penetration ​ | 
-|  640  |  E.g: Cy5, TexasRed, AL647  |  High penetration ​ | +|  ​<fc #ff2100>640 nm</​fc> ​ ​| ​ E.g: Cy5, TexasRed, AL647  |  High penetration ​ | 
-|  785  |  E.g: Atto740, Atto790 ​ |  Highest penetration ​ |+|  ​<fc #be0000>785 nm</​fc> ​ ​| ​ E.g: Atto740, Atto790 ​ |  Highest penetration ​ |
  
 ===== Fluorescence filters ===== ===== Fluorescence filters =====
  
-^  Position ​ ^  Filterset ​ ^  Emission ​(nm)  ^ +^  Position ​ ^  Filterset ​ ^  Emission ​ ^ 
-|  1  |  488  |  ​525/50  | +|  1  |  488  |  ​<fc #​4aff00>​500-550 nm</fc> ​ | 
-|  2  |  561  |  ​595/40  | +|  2  |  561  |  ​<fc #​ffcf00>​575-615 nm</fc> ​ | 
-|  3  |  640  |  ​680/30  | +|  3  |  640  |  ​<fc #​ff0000>​665-695 nm</fc> ​ | 
-|  4  |  785  |  ​LP 805  |+|  4  |  785  |  ​<fc #be0000> >805 nm</​fc> ​ |
 |  5  |  Empty  |  -  | |  5  |  Empty  |  -  |
 |  6  |  Empty  |  -  | |  6  |  Empty  |  -  |
Line 55: Line 55:
 ===== Objectives ===== ===== Objectives =====
 ^  Magnification ​ ^  Dipping caps RI range ^  NA  ^  WD per dipping cap (mm)  ^  Maximum Theoretical Resolution (um)  ^ ^  Magnification ​ ^  Dipping caps RI range ^  NA  ^  WD per dipping cap (mm)  ^  Maximum Theoretical Resolution (um)  ^
-|  1.1x  |  1.33-1.49 / 1.50-1.57 ​ |  0.1  |  16 / 17.6  |  4.8  | +|  ​<fc #88888a>1.1x</​fc> ​ ​| ​ 1.33-1.49 / 1.50-1.57 ​ |  0.1  |  16 / 17.6  |  4.8  | 
-|  4x  |  1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.35  |  >=15 / >=15 / >​=15 ​ |  1.3  | +|  ​<fc #df2b37>4x</​fc> ​ ​| ​ 1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.35  |  >=15 / >=15 / >​=15 ​ |  1.3  | 
-|  12x  |  1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.55  |  8.5 / 10.9 / 10  |  0.5  |+|  ​<fc #fef042>12x</​fc> ​ ​| ​ 1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.55  |  8.5 / 10.9 / 10  |  0.5  |
  
 ===== Objectives + Zoom ===== ===== Objectives + Zoom =====
-^  Objective ​ ^  Zoom  ^  Final Magnification ​ ​^ ​ Field of view (mm)  ^ +^  Objective ​ ^  Zoom  ^  Final Magnification ​ ^ 
-|  1.1x  |  0.6x  |  0.66x  ​|    ​+|  ​<fc #88888a>1.1x</​fc> ​ ​| ​ 0.6x  |  0.66x  | 
-|  1.1x  |  1x  |  1.1x  ​|    ​+|  ​<fc #88888a>1.1x</​fc> ​ ​| ​ 1x  |  1.1x  | 
-|  1.1x  |  1.66x  |  ~1.83x  ​|    ​+|  ​<fc #88888a>1.1x</​fc> ​ ​| ​ 1.66x  |  ~1.83x ​ | 
-|  1.1x  |  2.5x  |  2.75x  ​|    ​+|  ​<fc #88888a>1.1x</​fc> ​ ​| ​ 2.5x  |  2.75x  | 
-|  4x  |  0.6x  |  2.4x  ​|    ​+|  ​<fc #df2b37>4x</​fc> ​ ​| ​ 0.6x  |  2.4x  | 
-|  4x  |  1x  |  4x  ​|    ​+|  ​<fc #df2b37>4x</​fc> ​ ​| ​ 1x  |  4x  | 
-|  4x  |  1.66x  |  6.64x  ​|    ​+|  ​<fc #df2b37>4x</​fc> ​ ​| ​ 1.66x  |  6.64x  | 
-|  4x  |  2.5x  |  10x  ​|    ​+|  ​<fc #df2b37>4x</​fc> ​ ​| ​ 2.5x  |  10x  | 
-|  12x  |  0.6x  |  7.2x  ​|    ​+|  ​<fc #fef042>12x</​fc> ​ ​| ​ 0.6x  |  7.2x  | 
-|  12x  |  1x  |  12x  ​|    ​+|  ​<fc #fef042>12x</​fc> ​ ​| ​ 1x  |  12x  | 
-|  12x  |  1.66x  |  19.92x  ​|    ​+|  ​<fc #fef042>12x</​fc> ​ ​| ​ 1.66x  |  19.92x ​ | 
-|  12x  |  2.5x  |  30x  ​|    ​|+|  ​<fc #fef042>12x</​fc> ​ ​| ​ 2.5x  |  30x  |
  
 ===== Camera ===== ===== Camera =====
Line 82: Line 82:
 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
   * Turn on the power bar on **top of the microscope'​s computer**, under the table.   * Turn on the power bar on **top of the microscope'​s computer**, under the table.
 +  * 
 +  * Start the workstation. Booting will take a long time, and will only beep and turn the "​ON"​ light after 2-3min - be patient! ​
  
 ===== Microscope Turn Off Procedure ===== ===== Microscope Turn Off Procedure =====
blaze.1771344192.txt.gz · Last modified: 2026/02/17 17:03 by bioimaging