blaze [2025/01/13 13:25]
bioimaging
blaze [2025/02/03 12:51] (current)
bioimaging
Line 13: Line 13:
 <​b>​Year</​b>:​ 2024 <br> <​b>​Year</​b>:​ 2024 <br>
 <​b>​SN</​b>: ​ <br> <​b>​SN</​b>: ​ <br>
- 
-&nbsp <br> 
 &nbsp <br> &nbsp <br>
 &rarr; <a href="/​facility/​bioimaging/​doku.php?​id=blaze_quality"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=list-icon.png"​ width=18>​ Miltenyi UltraMicroscope Blaze Quality Control</​a>​ &rarr; <a href="/​facility/​bioimaging/​doku.php?​id=blaze_quality"><​img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=list-icon.png"​ width=18>​ Miltenyi UltraMicroscope Blaze Quality Control</​a>​
Line 30: Line 28:
 Equipped with three objectives (1.1x, 4x and 12x) and internal magnifications of 0.6x, 1x, 1.66x and 2.5x, this system is capable of going from whole organs to single cell level. Equipped with three objectives (1.1x, 4x and 12x) and internal magnifications of 0.6x, 1x, 1.66x and 2.5x, this system is capable of going from whole organs to single cell level.
 There are 4 laser lines in the Blaze, 488, 561, 640 and 785 nm, and with the capability of combining these with any of the installed filters, the system is capable of mix-matching for markers such as Propidium Iodide. There are 4 laser lines in the Blaze, 488, 561, 640 and 785 nm, and with the capability of combining these with any of the installed filters, the system is capable of mix-matching for markers such as Propidium Iodide.
 +
 +===== Usage conditions =====
 +
 +The **default configuration** for the UltraMicroscope Blaze is for the **Refractive Index of DBE** (iDisco protocol) at **RI = 1.56**. ​
 +Any use where another immersion media is necessary, the system will need realignment,​ and possibly undergo objective dipping-cap changes (below RI = 1.49). The **time taken to do these alignments** (DBE to immersion media, immersion media to DBE) can vary between 1-3 hours, and **will be charged to the user** in question.
  
 ===== Laser Lines ===== ===== Laser Lines =====
Line 46: Line 49:
 |  3  |  640  |  680/​30 ​ | |  3  |  640  |  680/​30 ​ |
 |  4  |  785  |  LP 805  | |  4  |  785  |  LP 805  |
-|  5  |  ​ ​|  ​Empty  | +|  5  |  ​Empty  ​|  ​ | 
-|  6  |  -  |  Empty  |+|  6  ​| ​ Empty  ​| ​ -  | 
 +|  7  ​| ​ Empty  ​| ​ -  |
  
 ===== Objectives ===== ===== Objectives =====
-^  Magnification ​ ^  Dipping caps RI range ^  NA  ^  WD per dipping cap (mm)  ^  Maximum Theoretical Resolution (um)+^  Magnification ​ ^  Dipping caps RI range ^  NA  ^  WD per dipping cap (mm)  ^  Maximum Theoretical Resolution (um)  ^
 |  1.1x  |  1.33-1.49 / 1.50-1.57 ​ |  0.1  |  16 / 17.6  |  4.8  | |  1.1x  |  1.33-1.49 / 1.50-1.57 ​ |  0.1  |  16 / 17.6  |  4.8  |
 |  4x  |  1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.35  |  >=15 / >=15 / >​=15 ​ |  1.3  | |  4x  |  1.33-1.41 / 1.42-1.48 / 1.49-1.57 ​ |  0.35  |  >=15 / >=15 / >​=15 ​ |  1.3  |
Line 57: Line 61:
 ===== Objectives + Zoom ===== ===== Objectives + Zoom =====
 ^  Objective ​ ^  Zoom  ^  Final Magnification ​ ^  Field of view (mm)  ^ ^  Objective ​ ^  Zoom  ^  Final Magnification ​ ^  Field of view (mm)  ^
-|  1.1x  |  0.6x  |  0.66x  |  ​33  ​|+|  1.1x  |  0.6x  |  0.66x  |    |
 |  1.1x  |  1x  |  1.1x  |    | |  1.1x  |  1x  |  1.1x  |    |
 |  1.1x  |  1.66x  |  ~1.83x ​ |    | |  1.1x  |  1.66x  |  ~1.83x ​ |    |
-|  1.1x  |  2.5x  |  2.75x  |  ​7.9  ​+|  1.1x  |  2.5x  |  2.75x  |    
-|  4x  |  0.6x  |  2.4x  |  ​9.1  ​|+|  4x  |  0.6x  |  2.4x  |    |
 |  4x  |  1x  |  4x  |    | |  4x  |  1x  |  4x  |    |
 |  4x  |  1.66x  |  6.64x  |    | |  4x  |  1.66x  |  6.64x  |    |
-|  4x  |  2.5x  |  10x  |  ​2.2  ​+|  4x  |  2.5x  |  10x  |    
-|  12x  |  0.6x  |  7.2x  |  ​3  ​|+|  12x  |  0.6x  |  7.2x  |    |
 |  12x  |  1x  |  12x  |    | |  12x  |  1x  |  12x  |    |
 |  12x  |  1.66x  |  19.92x ​ |    | |  12x  |  1.66x  |  19.92x ​ |    |
-|  12x  |  2.5x  |  30x  |  ​0.73  ​|+|  12x  |  2.5x  |  30x  |    |
  
 ===== Camera ===== ===== Camera =====
  
 +^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^
 +|  [[https://​www.excelitas.com/​product/​pcoedge-55-usb-scmos-camera|PCO Edge 5.5 sCMOS]] ​ |  2048 × 2048  |  6.5 x 6.5  |  60%  |
 +
 +
 +===== Microscope Turn On Procedure =====
 +  * Turn on the power bar on **top of the microscope'​s computer**, under the table.
 +
 +===== Microscope Turn Off Procedure =====
 +  * With the software still open, **pull out the cuvette drawer**.
 +  * Wait for the **drawer** to move **all the way down**.
 +  * **Remove any used cuvette**.
 +  * **Press the drawer button**, but **do not push the drawer** in.
 +  * In software, **change to 12x objective** and **lower the objectives all the way**.
 +  * Turn off the power bar on **top of the microscope'​s computer**, under the table.
 +  * **Close** the cuvette **drawer and software**.
  
 ---- ----
 [[resources|Back to the Equipment page]] [[resources|Back to the Equipment page]]