This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
agilent_cytation_5 [2025/02/03 11:19] bioimaging |
agilent_cytation_5 [2025/07/19 21:27] (current) bioimaging |
||
---|---|---|---|
Line 12: | Line 12: | ||
<b>Software</b>: Gen 5 Image Prime <br> | <b>Software</b>: Gen 5 Image Prime <br> | ||
<b>Year</b>: 2024 <br> | <b>Year</b>: 2024 <br> | ||
- | <b>SN</b>: <br> | + | <b>SN</b>: JDGD GHWD HGWM KH<br> |
  <br> |   <br> | ||
Data will be deleted after: <b>1 month</b> | Data will be deleted after: <b>1 month</b> | ||
Line 62: | Line 62: | ||
Images coming from the Cytation 5 are not calibrated, even though the system has that information available. In the images' metadata you can find the PixelWidth and the ImageWidthMicrons, and by dividing the latter for the first, you get the Pixel Size to calibrate your images. | Images coming from the Cytation 5 are not calibrated, even though the system has that information available. In the images' metadata you can find the PixelWidth and the ImageWidthMicrons, and by dividing the latter for the first, you get the Pixel Size to calibrate your images. | ||
- | Below you can also find this information, per objective and Field Of View (FOV): | + | You can find this information below, per objective and Field Of View (FOV): |
^ Objective ^ FOV ^ Image Size (pixels) ^ Image Size (microns) ^ Pixel Size (microns) ^ | ^ Objective ^ FOV ^ Image Size (pixels) ^ Image Size (microns) ^ Pixel Size (microns) ^ | ||
Line 90: | Line 90: | ||
* Turn on the microscope using the **switch** at the **front of the system**. | * Turn on the microscope using the **switch** at the **front of the system**. | ||
{{cytation_turnon_1.png?200|}} | {{cytation_turnon_1.png?200|}} | ||
- | * Turn the **computer** on. | + | * Turn the **computer** on (if needed). |
{{cytation_turnon_2.png?200|}} | {{cytation_turnon_2.png?200|}} | ||
* **Login** to Windows with your **GIMM Agendo credentials**. | * **Login** to Windows with your **GIMM Agendo credentials**. | ||
Line 103: | Line 103: | ||
* Turn off the microscope using the **switch** at the **front of the system**. | * Turn off the microscope using the **switch** at the **front of the system**. | ||
{{cytation_turnon_1.png?200|}} | {{cytation_turnon_1.png?200|}} | ||
- | * **Sign out** of the computer. | + | * **Log off** the computer. |
* **If you used the gas controller** | * **If you used the gas controller** |