Line 12: Line 12:
 <​b>​Software</​b>:​ Gen 5 Image Prime <br> <​b>​Software</​b>:​ Gen 5 Image Prime <br>
 <​b>​Year</​b>:​ 2024 <br> <​b>​Year</​b>:​ 2024 <br>
-<​b>​SN</​b>: ​ <br>+<​b>​SN</​b>: ​JDGD GHWD HGWM KH<br>
 &nbsp <br> &nbsp <br>
 Data will be deleted after: <b>1 month</​b>​ Data will be deleted after: <b>1 month</​b>​
Line 37: Line 37:
 ===== Filtersets available ===== ===== Filtersets available =====
  
-^  Filterset ​ ^  Filterset Reference ​ ^  LED wavelength (nm)  ^  LED reference ​ ^  Excitation (nm)  ^  Dichroic (nm)  ^  Emission (nm)  ^  ​Default Config ​ +^  Default Config  ​^  Filterset ​ ^  Filterset Reference ​ ^  LED wavelength (nm)  ^  LED reference ​ ^  Excitation (nm)  ^  Dichroic (nm)  ^  Emission (nm)  ^ 
-|  Laser Autofocus ​ |  [[https://​www.agilent.com/​store/​productDetail.jsp?​catalogId=1225010&​catId=SubCat2ECS_2041516|1225010]] ​ |  -  |  -  |  -  |  -  |  -  |  Pos  + Pos 1  ​| ​ Laser Autofocus ​ |  [[https://​www.agilent.com/​store/​productDetail.jsp?​catalogId=1225010&​catId=SubCat2ECS_2041516|1225010]] ​ |  -  |  -  |  -  |  -  |  -  ​
-|  DAPI  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225100/​1225100|1225100]] ​ |  365  |  1225007 ​ |  377/​50 ​ |  409  |  447/​60 ​ |  ​Pos 2  +|  Pos  ​| ​ DAPI  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225100/​1225100|1225100]] ​ |  365  |  1225007 ​ |  377/​50 ​ |  409  |  447/​60  ​
-|  GFP  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225101/​1225101|1225101]] ​ |  465  |  1225001 ​ |  469/​35 ​ |  497  |  525/​39 ​ |  ​ +|  ​ ​| ​ GFP  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225101/​1225101|1225101]] ​ |  465  |  1225001 ​ |  469/​35 ​ |  497  |  525/​39  ​
-|  Propidium Iodide ​ |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225111/​1225111|1225111]] ​ |  523  |  1225003 ​ |  531/​40 ​ |  605  |  647/​57 ​ |  ​Pos 3  +|  ​Pos 3  ​| ​ Propidium Iodide ​ |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225111/​1225111|1225111]] ​ |  523  |  1225003 ​ |  531/​40 ​ |  605  |  647/​57  ​
-|  Texas Red  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225102/​1225102|1225102]] ​ |  590  |  1225002 ​ |  586/​15 ​ |  605  |  647/​57 ​ |  ​ +|  ​ ​| ​ Texas Red  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225102/​1225102|1225102]] ​ |  590  |  1225002 ​ |  586/​15 ​ |  605  |  647/​57  ​
-|  Cy5  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225105/​|1225105]] ​ |  623  |  1225005 ​ |  628/​40 ​ |  660  |  685/​40 ​ ​| ​ Pos 4  |+|  ​Pos 4  ​| ​ Cy5  |  [[https://​www.agilent.com/​store/​en_US/​Prod-1225105/​|1225105]] ​ |  623  |  1225005 ​ |  628/​40 ​ |  660  |  685/​40 ​ |
  
  
Line 59: Line 59:
  
  
-===== Microscope Turn On Procedure ​=====+===== Pixel Size =====
  
-  * Turn on the microscope using the **switch** at the **front of the system**.+Images coming from the Cytation 5 are not calibrated, even though ​the system has that information available. In the images'​ metadata you can find the PixelWidth and the ImageWidthMicrons,​ and by dividing the latter for the first, you get the Pixel Size to calibrate your images. 
 +You can find this information below, per objective and Field Of View (FOV):
  
-  * **If you require CO2/O2 control**, turn the **atmosphere controller ON**+^  Objective ​ ^  FOV  ^  Image Size (pixels) ​ ^  Image Size (microns) ​ ^  Pixel Size (microns) ​ ^ 
 +|  4x  |  Full  |  1992 x 1992  |  3474 x 3474  |  ~1.7440 x ~1.7440 ​ | 
 +|  4x  |  75%  |  1496 x 1496  |  2609 x 2609  |  ~1.7440 x ~1.7440 ​ | 
 +|  4x  |  Standard ​ |  1224 x 904  |  2135 x 1576  |  ~1.7443 x ~1.7434 ​ | 
 +|  20x  |  Full  |  1992 x 1992  |  694 x 694  |  ~0.3484 x ~0.3484 ​ | 
 +|  20x  |  75%  |  1496 x 1496  |  521 x 521  |  ~0.3483 x ~0.3483 ​ | 
 +|  20x  |  Standard ​ |  1224 x 904  |  427 x 315  |  ~0.3489 x ~0.3485 ​ | 
 +|  40x  |  Full  |  1992 x 1992  |  347 x 347  |  ~0.1742 x ~0.1742 ​ | 
 +|  40x  |  75%  |  1496 x 1496  |  260 x 260  |  ~0.1738 x ~0.1738 ​ | 
 +|  40x  |  Standard ​ |  1224 x 904  |  213 x 157  |  ~0.1740 x ~0.1737 ​ |
  
-* Confirm the **CO2/O2 switch is ON**, set to the desired percentage, and **open the CO2/N2 tap** on the wall. 
  
-  * Turn the **computer** on.+===== CO2 and/or O2 Control Procedure ​ =====
  
-  * **Login** to Windows with your **GIMM Agendo credentials**.+  * **If you require CO2 and/or O2 control**, turn the **atmosphere controller ON**. 
 +{{cytation_gas_1.png?​200|}} 
 +  * Confirm the **CO2 and/or O2 switch is ON**, and set to the desired percentage. 
 +{{cytation_gas_2.png?​600|}} 
 +  ​* **Open the CO2 and/or N2 tap** on the wall. 
 +{{cytation_gas_3.png?200|}}
  
-  * Wait for the **button light** in the front to **turn green**. 
  
 +===== Microscope Turn On Procedure =====
 +
 +  * Turn on the microscope using the **switch** at the **front of the system**.
 +{{cytation_turnon_1.png?​200|}}
 +  * Turn the **computer** on.
 +{{cytation_turnon_2.png?​200|}}
 +  * **Login** to Windows with your **GIMM Agendo credentials**.
 +
 +  * Wait for the **button light** in the front to **turn green** (the system will pop out the plate drawer as well).
 +{{cytation_turnon_3.png?​200|}}
   * Open the **Gen 5** software.   * Open the **Gen 5** software.
  
 +
 +===== Microscope Turn Off Procedure =====
 +
 +  * Turn off the microscope using the **switch** at the **front of the system**.
 +{{cytation_turnon_1.png?​200|}}
 +  * **Sign out** of the computer.
 +
 +  * **If you used the gas controller** ​
 +    * Turn gas controller OFF
 +{{cytation_gas_1.png?​200|}}
 +    * Close all wall taps.
 +{{cytation_gas_3.png?​200|}}
  
 ---- ----
 [[resources|Back to the Equipment page]] [[resources|Back to the Equipment page]]