<b>SN</b>: 2841000239 <br> | <b>SN</b>: 2841000239 <br> | ||
  <br> |   <br> | ||
+ | Data will be deleted after: <b>1 month</b> | ||
  <br> |   <br> | ||
  <br> |   <br> | ||
Line 62: | Line 63: | ||
==== LEDs (Ocular) ==== | ==== LEDs (Ocular) ==== | ||
- | ^ Type ^ Wavelength ^ | + | ^ Line ^ Wavelength ^ |
| UV | 405 nm | | | UV | 405 nm | | ||
- | | Blue | 475 nm | | + | | B | 475 nm | |
- | | Green | 555 nm | | + | | G | 555 nm | |
- | | Red | 630 nm | | + | | R | 630 nm | |
==== Filterset (Ocular) ==== | ==== Filterset (Ocular) ==== | ||
Line 77: | Line 78: | ||
===== Microscope Turn On Procedure ===== | ===== Microscope Turn On Procedure ===== | ||
+ | |||
+ | The system should be always on, you should only need to turn on the computer, log into Windows and start **ZEN**. | ||
+ | |||
+ | If the system is completely off (e.g. during weekends), please follow this procedure: | ||
+ | |||
* Turn on the ''MAIN SWITCH'' | * Turn on the ''MAIN SWITCH'' | ||
{{::zeiss_lsm_980_main_switch.png?100|}} | {{::zeiss_lsm_980_main_switch.png?100|}} | ||
Line 82: | Line 88: | ||
{{::zeiss_lsm_980_laser_key.png?100|}} | {{::zeiss_lsm_980_laser_key.png?100|}} | ||
* Turn on the computer | * Turn on the computer | ||
- | * Login in to Windows with your Agendo credentials | + | * Log into Windows with your Agendo credentials |
* Turn on the ''COMPONENTS'' switch | * Turn on the ''COMPONENTS'' switch | ||
{{::zeiss_lsm_980_components.png?100|}} | {{::zeiss_lsm_980_components.png?100|}} | ||
Line 88: | Line 94: | ||
===== Microscope Turn Off Procedure ===== | ===== Microscope Turn Off Procedure ===== | ||
- | If there is another user for this microscope in the __next hour__: | + | |
- | * Exit the **ZEN** software | + | Please do not shutdown the system. Just exit the **ZEN** software, log off Windows and shutdown the computer. |
- | * Log off the computer | + | Don't forget to clean immersion objectives! |
- | * Clean up immersion objectives | + | |
- | Else: | + | |
- | * Turn off the lasers on **ZEN** | + | |
- | * Clean up immersion objectives | + | |
- | * Exit the **ZEN** software | + | |
- | * Shutdown the computer | + | |
- | * Turn off the ''COMPONENTS'' switch | + | |
- | * Turn off the ''Laser'' key | + | |
- | * Turn off the ''MAIN SWITCH'' | + | |
---- | ---- | ||
[[resources|Back to the Equipment page]] | [[resources|Back to the Equipment page]] |