{{:info.png?25|}} Click on the image on the right to see the system beam path in higher detail | {{:info.png?25|}} Click on the image on the right to see the system beam path in higher detail | ||
- | {{:warning.png?25|}} Data files older than **1 month** will be automatically deleted on this system, please copy your data to the iMM server using the desktop link. | + | {{:warning.png?25|}} Data files older than **3 months** will be automatically deleted on this system, please copy your data to the iMM server using the desktop link. |
===== Booking Rules ===== | ===== Booking Rules ===== | ||
Line 76: | Line 76: | ||
* Turn on the two switches: ''SYSTEMS/PC'' and ''COMPONENTS'' | * Turn on the two switches: ''SYSTEMS/PC'' and ''COMPONENTS'' | ||
{{switches_710.jpg?100}} | {{switches_710.jpg?100}} | ||
- | * Turn the metal halide power source on (if needed) | + | * Turn on the metal halide fluorescent light source (if needed) |
{{halide_710.jpg?80}} | {{halide_710.jpg?80}} | ||
- | * Turn the Ar laser on using the key (if needed) | + | * Turn on the Ar laser using the key (if needed) |
{{laser_key_710.jpg?100}} | {{laser_key_710.jpg?100}} | ||
- | * Turn the computer on | + | * Turn on the computer |
- | * Login with your unit account | + | * Log in to Windows (Bioimaging User) |
- | * Wait for network icon to change to ''connected'' state | + | * Login with Bioimaging User account |
+ | * Check that the computer is connected to an ''Unidentified network'' (click on the network icon in the taskbar) | ||
+ | {{::network_icon.png?200}} | ||
* Switch the Ar laser from ''idle'' ton ''run'' (don't change the ''current'') | * Switch the Ar laser from ''idle'' ton ''run'' (don't change the ''current'') | ||
{{laser_run_710.jpg?100}} | {{laser_run_710.jpg?100}} | ||
- | * Start the **ZEN 2012** software | + | * Start the **ZEN** software |
===== Microscope Turn Off Procedure ===== | ===== Microscope Turn Off Procedure ===== | ||
- | If there's another user for this microscope in the next hour: | + | If there is another user for this microscope in the __next hour__: |
- | * Close the software, leave the lasers on and log off the computer. | + | * Close **ZEN**, leave the lasers on, log off the computer |
- | * Clean up any objective where you used immersion oil. | + | * Clean up immersion objectives |
- | * **Make sure that there really is another user going to use the microscope.** | + | |
Else: | Else: | ||
* Turn off the 561 nm laser and incubator in the software (if used) | * Turn off the 561 nm laser and incubator in the software (if used) | ||
* Switch the Ar laser from ''run'' to ''idle'' (if used) | * Switch the Ar laser from ''run'' to ''idle'' (if used) | ||
* Turn off the Ar laser using the key (if used) | * Turn off the Ar laser using the key (if used) | ||
- | * Clean up any objective were you used immersion oil. | + | * Clean up immersion objectives |
- | * Exit the software, shut down the computer. | + | * Close **ZEN**, shutdown the computer. |
- | * Turn off the metal halide lamp | + | * Turn off the metal halide fluorescence light source |
- | * Turn off the main switches ''SYSTEMS/PC'' and ''COMPONENTS'' | + | * Turn off the two switches: ''SYSTEMS/PC'' and ''COMPONENTS'' |
* Wait 5 min. for Ar laser cooldown | * Wait 5 min. for Ar laser cooldown | ||
- | * Turn off the ''MAIN SWITCH''. | + | * Turn off the ''MAIN SWITCH'' |
---- | ---- | ||
[[resources|Back to the Resources page]] | [[resources|Back to the Resources page]] |