 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
-  * Turn on the computer+  ​* Turn on the HXP 200 fluorescent light source (if used) 
 +  ​* Turn on the computer ​(underneath the microscope table)
   * Login in to Windows (Bioimaging User)   * Login in to Windows (Bioimaging User)
-  * Turn on the black socket outlet **number 1** underneath the microscope table+  * Turn on the black socket outlet **number 1** (back left corner of microscope table)
 {{::​sd_power.png?​200|}} {{::​sd_power.png?​200|}}
   * Wait for the Definite Focus controller to show the App-Ver. and OS-Image info   * Wait for the Definite Focus controller to show the App-Ver. and OS-Image info
 {{::​df_on.png?​500|}} {{::​df_on.png?​500|}}
-  * Turn on the black socket outlet **number 2** underneath the microscope table +  * Turn on the black socket outlet **number 2** 
-  * Turn on the black socket outlet **number 3** underneath the microscope table (for spinning disk mode)+  * Turn on the black socket outlet **number 3**
   * Start the **ZEN Blue** software   * Start the **ZEN Blue** software
- 
-**Warning**:​ If you need to change the stage adapter, please contact the <​html><​a href="/​facility/​bioimaging/​doku.php?​id=about">​Bioimaging Unit</​a>​ (<​b>​imm-bioimaging@medicina.ulisboa.pt</​b>​ | <img src="/​facility/​bioimaging/​lib/​exe/​fetch.php?​media=phone_neg.png"​ width=15>​ 47222</​html>​) 
- 
 ===== Microscope Turn Off Procedure ===== ===== Microscope Turn Off Procedure =====
 If there is another user for this microscope in the __next hour__: If there is another user for this microscope in the __next hour__:
Line 123: Line 121:
 Else: Else:
   * Exit the **ZEN** software   * Exit the **ZEN** software
-  * Shut down the computer+  * Shut down the computer ​(not the UPS!)
   * Clean up immersion objectives   * Clean up immersion objectives
-  * Turn off the black socket outlets ​underneath the microscope table by reverse order (3→2→1)+  * Turn off the black socket outlets by reverse order (3→2→1) 
 +  * Turn off the HXP 200 fluorescent light source
  
 ---- ----
  
 [[resources|Back to the Equipment page]] [[resources|Back to the Equipment page]]