leica_sp8_mp [2020/01/20 16:44] bioimaging [System Turn On Procedures] |
leica_sp8_mp [2020/02/28 00:36] bioimaging [LASERs] |
||
---|---|---|---|
Line 39: | Line 39: | ||
==== LASERs ==== | ==== LASERs ==== | ||
^ Laser Unit ^ Wavelength ^ Maximum Power ^ Current Status ^ | ^ Laser Unit ^ Wavelength ^ Maximum Power ^ Current Status ^ | ||
- | | InSight DS+ Dual (multi-photon) | 680 - 1300 nm | 1.3 W | **not working** | | + | | InSight DS+ Dual (multi-photon) | 680 - 1300 nm | 1.3 W | ok | |
- | | ::: | 1041 nm | 1.5 W | **not working** | | + | | ::: | 1040 nm | 1.5 W | **not working** | |
| SS OBIS 488-20 (confocal) | 488 nm | 20 mW | ok | | | SS OBIS 488-20 (confocal) | 488 nm | 20 mW | ok | | ||
Line 97: | Line 97: | ||
* Turn on the microscope's control panel | * Turn on the microscope's control panel | ||
{{::4_control_panel.png?150|}} | {{::4_control_panel.png?150|}} | ||
- | * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser | + | * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser in the flexible supply unit |
{{::5_scanners_laser.png?120|}} | {{::5_scanners_laser.png?120|}} | ||
* Log in to Windows (Bioimaging User) | * Log in to Windows (Bioimaging User) | ||
* Start the **LAS X** software | * Start the **LAS X** software | ||
+ | |||
+ | ===== Microscope Turn Off procedures ===== | ||
+ | |||
+ | If there is another user for this microscope in the __next hour__: | ||
+ | * Leave the fluorescent lamp and lasers on | ||
+ | * Clean up immersion objectives | ||
+ | Else: | ||
+ | * Clean up immersion objectives | ||
+ | * Set the infrared laser to ''standby'' (deactivate the infrared laser in the Currently Available Lasers window) | ||
+ | * Close the **LAS X** software | ||
+ | * Turn of the 488 nm laser with the key switch in the flexible supply unit | ||
+ | * Switch off the HyD supply unit | ||
+ | * Switch off the microscope's control panel | ||
+ | * Shutdown the computer | ||
+ | * Switch off the laser power and scanner power in the flexible supply unit | ||
+ | * Turn off the microscope electronics box | ||
+ | * Turn off the fluorescent lamp | ||
---- | ---- | ||