Line 39: | Line 39: | ||
==== LASERs ==== | ==== LASERs ==== | ||
^ Laser Unit ^ Wavelength ^ Maximum Power ^ Current Status ^ | ^ Laser Unit ^ Wavelength ^ Maximum Power ^ Current Status ^ | ||
- | | InSight DS+ Dual (multi-photon) | 680 - 1300 nm | 1.3 W | **not working** | | + | | InSight DS+ Dual (multi-photon) | 680 - 1300 nm | 1.3 W | ok | |
- | | ::: | 1041 nm | 1.5 W | **not working** | | + | | ::: | 1040 nm | 1.5 W | **not working** | |
| SS OBIS 488-20 (confocal) | 488 nm | 20 mW | ok | | | SS OBIS 488-20 (confocal) | 488 nm | 20 mW | ok | | ||
==== Objectives ==== | ==== Objectives ==== | ||
^ Magnification ^ Model ^ Immersion ^ NA ^ WD (mm) ^ Reference ^ | ^ Magnification ^ Model ^ Immersion ^ NA ^ WD (mm) ^ Reference ^ | ||
- | | 25x | HC FLUOTAR VISIR | Water | 0.95 | 2.5 | 15506374 | | + | | 25x | HC FLUOTAR VISIR | Water | 0.95 | 2.50 | 15506374 | |
| 63x | HC PL APO Glyc CORR CS2 | Glycerol | 1.30 | 0.30 | 15506353 | | | 63x | HC PL APO Glyc CORR CS2 | Glycerol | 1.30 | 0.30 | 15506353 | | ||
Line 84: | Line 84: | ||
\\ | \\ | ||
\\ | \\ | ||
+ | |||
+ | ===== System Turn On Procedures ===== | ||
+ | * Check that the cooling and power supply for the infrared laser are ON (''InSight DeepSee READY'') | ||
+ | {{::0_infrared_laser.png?100|}} | ||
+ | * Turn on the fluorescence lamp | ||
+ | {{1_metal_halide.png?100|}} | ||
+ | * Turn on the HyD (Hybrid detectors) supply unit | ||
+ | {{::2_hyd_supply.png?100|}} | ||
+ | * Turn on the computer | ||
+ | * Turn on the microscope electronics box | ||
+ | {{::3_electronics_box.png?100|}} | ||
+ | * Turn on the microscope's control panel | ||
+ | {{::4_control_panel.png?150|}} | ||
+ | * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser in the flexible supply unit | ||
+ | {{::5_scanners_laser.png?120|}} | ||
+ | * Log in to Windows (Bioimaging User) | ||
+ | * Start the **LAS X** software | ||
+ | |||
+ | ===== Microscope Turn Off procedures ===== | ||
+ | |||
+ | If there is another user for this microscope in the __next hour__: | ||
+ | * Leave the fluorescent lamp and lasers on | ||
+ | * Clean up immersion objectives | ||
+ | Else: | ||
+ | * Clean up immersion objectives | ||
+ | * Set the infrared laser to ''hibernate'' (deactivate the infrared laser in the Currently Available Lasers window) | ||
+ | * Close the **LAS X** software | ||
+ | * Turn of the 488 nm laser with the key switch in the flexible supply unit | ||
+ | * Switch off the HyD supply unit | ||
+ | * Switch off the microscope's control panel | ||
+ | * Shutdown the computer | ||
+ | * Switch off the laser power and scanner power in the flexible supply unit | ||
+ | * Turn off the microscope electronics box | ||
+ | * Turn off the fluorescent lamp | ||
---- | ---- | ||