Line 70: | Line 70: | ||
\\ | \\ | ||
**Upon request:** | **Upon request:** | ||
- | (requires changing the dichroic beamsplitter) | + | \\ (requires changing filters and the dichroic beamsplitter) |
{{::alternative_configuration.png |}} | {{::alternative_configuration.png |}} | ||
Line 84: | Line 84: | ||
\\ | \\ | ||
\\ | \\ | ||
+ | |||
+ | ===== System Turn On Procedures ===== | ||
+ | * Check that the cooling and power supply for the infrared laser are ON (''InSight DeepSee READY'') | ||
+ | {{::0_infrared_laser.png?100|}} | ||
+ | * Turn on the fluorescence lamp | ||
+ | {{1_metal_halide.png?100|}} | ||
+ | * Turn on the HyD (Hybrid detectors) supply unit | ||
+ | {{::2_hyd_supply.png?100|}} | ||
+ | * Turn on the computer | ||
+ | * Turn on the microscope electronics box | ||
+ | {{::3_electronics_box.png?100|}} | ||
+ | * Turn on the microscope's control panel | ||
+ | {{::4_control_panel.png?150|}} | ||
+ | * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser | ||
+ | {{::5_scanners_laser.png?120|}} | ||
+ | * Log in to Windows (Bioimaging User) | ||
+ | * Start the **LAS X** software | ||
---- | ---- | ||