Line 55: Line 55:
 ==== Emission Filters (multi-photon) ==== ==== Emission Filters (multi-photon) ====
  
 +{{::​default_configuration.png ​ |}}
 +\\
 +\\
 +\\
 ^  Detector ​ ^  Filter ​ ^  Transmission ​ ^  Fluorochromes ​ ^ ^  Detector ​ ^  Filter ​ ^  Transmission ​ ^  Fluorochromes ​ ^
 |  HyD-RLD 1  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  | |  HyD-RLD 1  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  |
Line 60: Line 64:
 |  PMT-RLD 3  |  Far Red  |  662 - 737 nm  |  Alexa647, Cy5  | |  PMT-RLD 3  |  Far Red  |  662 - 737 nm  |  Alexa647, Cy5  |
 |  PMR-RLD 4  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  | |  PMR-RLD 4  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  |
- +\\ 
-{{::​default_configuration.png|}} +\\ 
 +\\ 
 +\\ 
 +\\
 **Upon request:** **Upon request:**
-(requires changing the dichroic beamsplitter)+\\ (requires changing ​filters and the dichroic beamsplitter)
  
 +{{::​alternative_configuration.png ​ |}}
 +\\
 +\\
 +\\
 ^  Detector ​ ^  Filter ​ ^  Transmission ​ ^  Fluorochromes ​ ^ ^  Detector ​ ^  Filter ​ ^  Transmission ​ ^  Fluorochromes ​ ^
 |  HyD-RLD 1  |  Green  |  500-550 nm  |  GFP, Alexa488, FITC  | |  HyD-RLD 1  |  Green  |  500-550 nm  |  GFP, Alexa488, FITC  |
Line 71: Line 81:
 |  PMT-RLD 3  |  Far Red  |  662 - 737 nm  |  Alexa647, Cy5  | |  PMT-RLD 3  |  Far Red  |  662 - 737 nm  |  Alexa647, Cy5  |
 |  PMR-RLD 4  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  | |  PMR-RLD 4  |  Red  |  570-640 nm  |  mCherry, Alexa568, Cy3  |
 +\\
 +\\
 +\\
  
-{{::alternative_configuration.png|}}+===== System Turn On Procedures ===== 
 +  * Check that the cooling and power supply for the infrared laser are ON (''​InSight DeepSee READY''​) 
 +{{::0_infrared_laser.png?100|}} 
 +  * Turn on the fluorescence lamp 
 +{{1_metal_halide.png?​100|}} 
 +  * Turn on the HyD (Hybrid detectors) supply unit 
 +{{::​2_hyd_supply.png?​100|}} 
 +  * Turn on the computer 
 +  * Turn on the microscope electronics box 
 +{{::​3_electronics_box.png?​100|}} 
 +  * Turn on the microscope'​s control panel 
 +{{::​4_control_panel.png?​150|}} 
 +  * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser 
 +{{::​5_scanners_laser.png?​120|}} 
 +  * Log in to Windows (Bioimaging User) 
 +  * Start the **LAS X** software
  
 ---- ----