This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
leica_sp8_mp [2020/01/17 13:10] bioimaging |
leica_sp8_mp [2020/01/20 16:44] bioimaging [System Turn On Procedures] |
||
---|---|---|---|
Line 55: | Line 55: | ||
==== Emission Filters (multi-photon) ==== | ==== Emission Filters (multi-photon) ==== | ||
+ | {{::default_configuration.png |}} | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | ^ Detector ^ Filter ^ Transmission ^ Fluorochromes ^ | ||
+ | | HyD-RLD 1 | Red | 570-640 nm | mCherry, Alexa568, Cy3 | | ||
+ | | HyD-RLD 2 | Green | 500-550 nm | GFP, Alexa488, FITC | | ||
+ | | PMT-RLD 3 | Far Red | 662 - 737 nm | Alexa647, Cy5 | | ||
+ | | PMR-RLD 4 | Red | 570-640 nm | mCherry, Alexa568, Cy3 | | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | **Upon request:** | ||
+ | \\ (requires changing filters and the dichroic beamsplitter) | ||
+ | |||
+ | {{::alternative_configuration.png |}} | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | ^ Detector ^ Filter ^ Transmission ^ Fluorochromes ^ | ||
+ | | HyD-RLD 1 | Green | 500-550 nm | GFP, Alexa488, FITC | | ||
+ | | HyD-RLD 2 | Blue | 415-485 nm | DAPI, Hoescht, Alexa350 | | ||
+ | | PMT-RLD 3 | Far Red | 662 - 737 nm | Alexa647, Cy5 | | ||
+ | | PMR-RLD 4 | Red | 570-640 nm | mCherry, Alexa568, Cy3 | | ||
+ | \\ | ||
+ | \\ | ||
+ | \\ | ||
+ | |||
+ | ===== System Turn On Procedures ===== | ||
+ | * Check that the cooling and power supply for the infrared laser are ON (''InSight DeepSee READY'') | ||
+ | {{::0_infrared_laser.png?100|}} | ||
+ | * Turn on the fluorescence lamp | ||
+ | {{1_metal_halide.png?100|}} | ||
+ | * Turn on the HyD (Hybrid detectors) supply unit | ||
+ | {{::2_hyd_supply.png?100|}} | ||
+ | * Turn on the computer | ||
+ | * Turn on the microscope electronics box | ||
+ | {{::3_electronics_box.png?100|}} | ||
+ | * Turn on the microscope's control panel | ||
+ | {{::4_control_panel.png?150|}} | ||
+ | * Switch on the scanner power, laser power and activate the key switch to turn on the 488 nm laser | ||
+ | {{::5_scanners_laser.png?120|}} | ||
+ | * Log in to Windows (Bioimaging User) | ||
+ | * Start the **LAS X** software | ||
---- | ---- | ||
[[resources|Back to the Resources page]] | [[resources|Back to the Resources page]] |