User Tools

Site Tools


3i_marianas_sdc

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
3i_marianas_sdc [2020/01/21 18:26]
bioimaging [Microscope Turn Off Procedure]
3i_marianas_sdc [2020/01/21 18:26]
bioimaging [Microscope Turn On Procedure]
Line 78: Line 78:
   * Turn on the fluorescence lamp (if needed)   * Turn on the fluorescence lamp (if needed)
 {{marianas_power5.jpg?​|}} {{marianas_power5.jpg?​|}}
-  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**+  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** ​from left to right
 {{marianas_power3.jpg?​|}} {{marianas_power3.jpg?​|}}
   * Turn on the secondary power supply switch (white socket outlet underneath the touchscreen)   * Turn on the secondary power supply switch (white socket outlet underneath the touchscreen)
3i_marianas_sdc.txt ยท Last modified: 2024/04/01 23:50 by bioimaging