3i_marianas_sdc [2020/01/21 18:25] bioimaging [Microscope Turn On Procedure] |
3i_marianas_sdc [2020/01/21 18:26] bioimaging [Microscope Turn On Procedure] |
||
---|---|---|---|
Line 78: | Line 78: | ||
* Turn on the fluorescence lamp (if needed) | * Turn on the fluorescence lamp (if needed) | ||
{{marianas_power5.jpg?|}} | {{marianas_power5.jpg?|}} | ||
- | * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** | + | * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** from left to right |
{{marianas_power3.jpg?|}} | {{marianas_power3.jpg?|}} | ||
* Turn on the secondary power supply switch (white socket outlet underneath the touchscreen) | * Turn on the secondary power supply switch (white socket outlet underneath the touchscreen) | ||
Line 101: | Line 101: | ||
* Switch all lasers off (first on the switch and finally by turning the key) | * Switch all lasers off (first on the switch and finally by turning the key) | ||
* Turn off the fluorescent lamp | * Turn off the fluorescent lamp | ||
+ | * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen) | ||
* | * | ||
* Turn off all system components (follow the Turn On Procedure in reverse) | * Turn off all system components (follow the Turn On Procedure in reverse) |