3i_marianas_sdc [2020/01/21 18:24]
bioimaging [Microscope Turn Off Procedure]
3i_marianas_sdc [2020/01/21 18:26]
bioimaging [Microscope Turn Off Procedure]
Line 80: Line 80:
   * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**   * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**
 {{marianas_power3.jpg?​|}} {{marianas_power3.jpg?​|}}
-  * Turn on the secondary power supply switch +  * Turn on the secondary power supply switch ​(white socket outlet underneath the touchscreen)
-{{marianas_power2.jpg?​|}}+
   * Turn on the Yokogawa spinning disk scanner (using the key)   * Turn on the Yokogawa spinning disk scanner (using the key)
 {{marianas_power4.jpg?​|}} {{marianas_power4.jpg?​|}}
Line 102: Line 101:
   * Switch all lasers off (first on the switch and finally by turning the key)   * Switch all lasers off (first on the switch and finally by turning the key)
   * Turn off the fluorescent lamp   * Turn off the fluorescent lamp
 +  * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen)
   *    * 
   * Turn off all system components (follow the Turn On Procedure in reverse)   * Turn off all system components (follow the Turn On Procedure in reverse)
3i_marianas_sdc.txt · Last modified: 2024/04/01 23:50 by bioimaging