3i_marianas_sdc [2020/01/01 18:12]
bioimaging [Objectives]
3i_marianas_sdc [2020/02/17 11:44]
bioimaging [Microscope Turn On Procedure]
Line 57: Line 57:
 |  40x  |  Plan-Apochromat Corr  |  Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ | |  40x  |  Plan-Apochromat Corr  |  Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ |
  
-==== Emission ​Filtersets ​====+==== Emission ​Filters  ​====
 ^  Setting ​ ^  Transmission ​ ^ ^  Setting ​ ^  Transmission ​ ^
 |  1: 445  |  422-477 nm  | |  1: 445  |  422-477 nm  |
Line 78: Line 78:
   * Turn on the fluorescence lamp (if needed)   * Turn on the fluorescence lamp (if needed)
 {{marianas_power5.jpg?​|}} {{marianas_power5.jpg?​|}}
-  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**+  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** ​from left to right
 {{marianas_power3.jpg?​|}} {{marianas_power3.jpg?​|}}
-  * Turn on the secondary power supply switch +  * Turn on the secondary power supply switch (white socket outlet underneath ​the touchscreen)
-{{marianas_power2.jpg?​|}} +
-  * Turn on the Yokogawa spinning disk scanner ​(using the key)+
 {{marianas_power4.jpg?​|}} {{marianas_power4.jpg?​|}}
   * Turn on incubation and CO2 on the touchscreen (if needed)   * Turn on incubation and CO2 on the touchscreen (if needed)
Line 98: Line 96:
   * Exit the **SlideBook** software   * Exit the **SlideBook** software
   * Switch off the computer   * Switch off the computer
-  * Turn off all system components ​(follow ​the Turn On Procedure in reverse)+  * Turn off incubation in the touchscreen (if used) 
 +  * Turn off CO2 (if used) 
 +  * Switch ​all lasers off (first on the switch and finally by turning the key) 
 +  * Turn off the fluorescent lamp 
 +  * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen) 
 +  * Turn off the component power switches (red) **one at a time** from right to left and the power switch distributor (green)
  
 ---- ----
  
 [[resources|Back to the Resources page]] [[resources|Back to the Resources page]]