 ==== Objectives ==== ==== Objectives ====
 ^  Magnification ​  ​^ ​ Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^ ^  Magnification ​  ​^ ​ Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^
-|  20x  |  Plan-Apochromat ​ |  Air  |  0.8  |  0.55  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420650-9902-000/​Objective-Plan-Apochromat-20x-0.8-M27#|420650-9902-000]] ​ |+|  10x  |  EC Plan-Neofluar ​ |  Air  |  0.30  |  5.5  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420340-9900-000/​Objective-EC-Plan-Neofluar-10x-0.30-M27|420340-9900-000]] ​ | 
 +|  20x  |  Plan-Apochromat ​ |  Air  |  0.8  |  0.55  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420650-9901-000/​Objective-Plan-Apochromat-20x-0.8-M27|420650-9901-000]] ​ | 
 +|  40x  |  LD C-Apochromat Corr  |  Water  |  1.1  |  0.62  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​421867-9970-000/​Objective-LD-C-Apochromat-40x-1.1-W-Corr-M27|421867-9970-000]] ​ |
 |  63x  |  Plan-Apochromat ​ |  Oil  |  1.40  |  0.19  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420780-9900-000/​Objective-Plan-Apochromat-63x-1.4-Oil-M27|420780-9900-000]] ​ | |  63x  |  Plan-Apochromat ​ |  Oil  |  1.40  |  0.19  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420780-9900-000/​Objective-Plan-Apochromat-63x-1.4-Oil-M27|420780-9900-000]] ​ |
 |  100x  |  Plan-Apochromat ​ |  Oil  |  1.40  |  0.17  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420790-9901-000/​Objective-Plan-Apochromat-100x-1.40-Oil-M27|420790-9901-000]] ​ | |  100x  |  Plan-Apochromat ​ |  Oil  |  1.40  |  0.17  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420790-9901-000/​Objective-Plan-Apochromat-100x-1.40-Oil-M27|420790-9901-000]] ​ |
Line 54: Line 56:
 **Upon request:** **Upon request:**
 ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^ ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^
-|  25x  |  LCI Plan-NeoFluar Imm Corr DIC  |  Oil/​Glyc/​W ​ |  0. ​| ​ 0.21  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420852-9973-000/​Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9973-000]] ​ |+|  25x  |  LCI Plan-Neofluar ​Corr DIC  |  Oil/​Glyc/​W ​ |  0.80  ​| ​ 0.21  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420852-9972-000/​Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9972-000]] ​ |
 |  40x  |  Plan-Apochromat Corr  |  Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ | |  40x  |  Plan-Apochromat Corr  |  Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ |
  
-==== Emission ​Filtersets ​====+==== Emission ​Filters  ​====
 ^  Setting ​ ^  Transmission ​ ^ ^  Setting ​ ^  Transmission ​ ^
 |  1: 445  |  422-477 nm  | |  1: 445  |  422-477 nm  |
Line 74: Line 76:
 |  [[https://​www.photometrics.com/​wp-content/​uploads/​2019/​10/​Evolve512-Datasheet.pdf|Photometrics Evolve 512 EMCCD]] ​ |  512 x 512  |  16 x 16  |  > 90%  | |  [[https://​www.photometrics.com/​wp-content/​uploads/​2019/​10/​Evolve512-Datasheet.pdf|Photometrics Evolve 512 EMCCD]] ​ |  512 x 512  |  16 x 16  |  > 90%  |
 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
-  * Check that the main power supply switch is on (should be on by default) 
-{{marianas_power1.jpg?​|}} 
   * Turn on the fluorescence lamp (if needed)   * Turn on the fluorescence lamp (if needed)
 {{marianas_power5.jpg?​|}} {{marianas_power5.jpg?​|}}
-  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**+  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** ​from left to right
 {{marianas_power3.jpg?​|}} {{marianas_power3.jpg?​|}}
-  * Turn on the secondary power supply switch +  * Turn on the secondary power supply switch (white socket outlet underneath ​the touchscreen)
-{{marianas_power2.jpg?​|}} +
-  * Turn on the Yokogawa spinning disk scanner ​(using the key) +
-{{marianas_power4.jpg?​|}}+
   * Turn on incubation and CO2 on the touchscreen (if needed)   * Turn on incubation and CO2 on the touchscreen (if needed)
 {{marianas_power6.jpg?​|}} {{marianas_power6.jpg?​|}}
Line 98: Line 95:
   * Exit the **SlideBook** software   * Exit the **SlideBook** software
   * Switch off the computer   * Switch off the computer
-  * Turn off all system components ​(follow ​the Turn On Procedure in reverse)+  * Turn off incubation in the touchscreen (if used) 
 +  * Turn off CO2 (if used) 
 +  * Switch ​all lasers off (first on the switch and finally by turning the key) 
 +  * Turn off the fluorescent lamp 
 +  * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen) 
 +  * Turn off the component power switches (red) **one at a time** from right to left and the power switch distributor (green)
  
 ---- ----
  