This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
3i_marianas_sdc [2020/01/01 17:31] bioimaging |
3i_marianas_sdc [2020/01/21 18:26] bioimaging [Microscope Turn On Procedure] |
||
---|---|---|---|
Line 6: | Line 6: | ||
<a href="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_wiki_new.png"><img src="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_wiki_new.png" width=300></a> | <a href="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_wiki_new.png"><img src="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_wiki_new.png" width=300></a> | ||
<td style="border:0px solid white;"> <p style="line-height:1.8"> | <td style="border:0px solid white;"> <p style="line-height:1.8"> | ||
- | <b>Location</b>: <a href="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_map_new.png">Room P2-A-22</a> (<img src="/facility/bioimaging/lib/exe/fetch.php?media=phone_neg.png" width=20> 47217) <br> | + | <b>Location</b>: <a href="/facility/bioimaging/lib/exe/fetch.php?media=3i_marianas_sdc_map_new.png">Room P2-A-22</a> (<img src="/facility/bioimaging/lib/exe/fetch.php?media=phone_neg.png" width=15> 47217) <br> |
<b>Manufacturer</b>: <a href="https://www.intelligent-imaging.com/" target="_blank">Intelligent Imaging Innovations</a><br> | <b>Manufacturer</b>: <a href="https://www.intelligent-imaging.com/" target="_blank">Intelligent Imaging Innovations</a><br> | ||
<b>Model</b>: <a href="https://www.intelligent-imaging.com/marianas" target="_blank">Marianas</a><br> | <b>Model</b>: <a href="https://www.intelligent-imaging.com/marianas" target="_blank">Marianas</a><br> | ||
Line 54: | Line 54: | ||
**Upon request:** | **Upon request:** | ||
^ Magnification ^ Model ^ Immersion ^ NA ^ WD (mm) ^ Reference ^ | ^ Magnification ^ Model ^ Immersion ^ NA ^ WD (mm) ^ Reference ^ | ||
- | | 25x | LCI Plan-NeoFluar Imm Corr DIC | Oil/Glyc/W | 0.8 | 0.21 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420852-9973-000/Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9973-000]] | | + | | 25x | LCI Plan-Neofluar Corr DIC | Oil/Glyc/W | 0.80 | 0.21 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420852-9972-000/Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9972-000]] | |
| 40x | Plan-Apochromat Corr | Air | 0.95 | 0.25 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420660-9970-000/Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] | | | 40x | Plan-Apochromat Corr | Air | 0.95 | 0.25 | [[https://www.micro-shop.zeiss.com/en/de/shop/objectives/420660-9970-000/Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] | | ||
- | ==== Emission Filtersets ==== | + | ==== Emission Filters ==== |
^ Setting ^ Transmission ^ | ^ Setting ^ Transmission ^ | ||
| 1: 445 | 422-477 nm | | | 1: 445 | 422-477 nm | | ||
Line 78: | Line 78: | ||
* Turn on the fluorescence lamp (if needed) | * Turn on the fluorescence lamp (if needed) | ||
{{marianas_power5.jpg?|}} | {{marianas_power5.jpg?|}} | ||
- | * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** | + | * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** from left to right |
{{marianas_power3.jpg?|}} | {{marianas_power3.jpg?|}} | ||
- | * Turn on the secondary power supply switch | + | * Turn on the secondary power supply switch (white socket outlet underneath the touchscreen) |
- | {{marianas_power2.jpg?|}} | + | |
* Turn on the Yokogawa spinning disk scanner (using the key) | * Turn on the Yokogawa spinning disk scanner (using the key) | ||
{{marianas_power4.jpg?|}} | {{marianas_power4.jpg?|}} | ||
Line 98: | Line 97: | ||
* Exit the **SlideBook** software | * Exit the **SlideBook** software | ||
* Switch off the computer | * Switch off the computer | ||
+ | * Turn off incubation in the touchscreen (if used) | ||
+ | * Turn off CO2 (if used) | ||
+ | * Switch all lasers off (first on the switch and finally by turning the key) | ||
+ | * Turn off the fluorescent lamp | ||
+ | * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen) | ||
+ | * | ||
* Turn off all system components (follow the Turn On Procedure in reverse) | * Turn off all system components (follow the Turn On Procedure in reverse) | ||