3i_marianas_sdc [2020/01/01 17:04]
bioimaging [System overview]
3i_marianas_sdc [2020/01/21 18:26]
bioimaging [Microscope Turn On Procedure]
Line 54: Line 54:
 **Upon request:** **Upon request:**
 ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^ ^  Magnification ​ ^  Model  ^  Immersion ​ ^  NA  ^  WD (mm)  ^  Reference ​ ^
-|  25x  |  LCI Plan-NeoFluar ​ ​| ​ Oil/​Glyc/​W ​ |  0. ​| ​ 0.21  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420852-9973-000/​Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9973-000]] ​ | +|  25x  |  LCI Plan-Neofluar Corr DIC  ​| ​ Oil/​Glyc/​W ​ |  0.80  ​| ​ 0.21  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420852-9972-000/​Objective-LCI-Plan-Neofluar-25x-0.8-Imm-Corr-DIC-M27|420852-9972-000]] ​ | 
-|  40x  |  Plan-Apochromat ​ |  Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ |+|  40x  |  Plan-Apochromat ​Corr  ​| ​ Air  |  0.95  |  0.25  |  [[https://​www.micro-shop.zeiss.com/​en/​de/​shop/​objectives/​420660-9970-000/​Objective-Plan-Apochromat-40x-0.95-Corr-M27|420660-9970-000]] ​ |
  
-==== Emission ​Filtersets ​====+==== Emission ​Filters  ​====
 ^  Setting ​ ^  Transmission ​ ^ ^  Setting ​ ^  Transmission ​ ^
 |  1: 445  |  422-477 nm  | |  1: 445  |  422-477 nm  |
Line 72: Line 72:
 ==== Camera ==== ==== Camera ====
 ^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^ ^  Model  ^  Frame Size  ^  Pixel Size (µm)  ^  Quantum Efficiency ​ ^
-|  Photometrics Evolve 512 EMCCD  |  512 x 512  |  16 x 16  |  > 90%  |+|  ​[[https://​www.photometrics.com/​wp-content/​uploads/​2019/​10/​Evolve512-Datasheet.pdf|Photometrics Evolve 512 EMCCD]]  ​| ​ 512 x 512  |  16 x 16  |  > 90%  |
 ===== Microscope Turn On Procedure ===== ===== Microscope Turn On Procedure =====
   * Check that the main power supply switch is on (should be on by default)   * Check that the main power supply switch is on (should be on by default)
Line 78: Line 78:
   * Turn on the fluorescence lamp (if needed)   * Turn on the fluorescence lamp (if needed)
 {{marianas_power5.jpg?​|}} {{marianas_power5.jpg?​|}}
-  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time**+  * Turn on the power switch distributor (green) and the component power switches (red) **one at a time** ​from left to right
 {{marianas_power3.jpg?​|}} {{marianas_power3.jpg?​|}}
-  * Turn on the secondary power supply switch +  * Turn on the secondary power supply switch ​(white socket outlet underneath the touchscreen)
-{{marianas_power2.jpg?​|}}+
   * Turn on the Yokogawa spinning disk scanner (using the key)   * Turn on the Yokogawa spinning disk scanner (using the key)
 {{marianas_power4.jpg?​|}} {{marianas_power4.jpg?​|}}
Line 98: Line 97:
   * Exit the **SlideBook** software   * Exit the **SlideBook** software
   * Switch off the computer   * Switch off the computer
 +  * Turn off incubation in the touchscreen (if used)
 +  * Turn off CO2 (if used)
 +  * Switch all lasers off (first on the switch and finally by turning the key)
 +  * Turn off the fluorescent lamp
 +  * Turn off the secondary power supply switch (white socket outlet underneath the touchscreen)
 +  * 
   * Turn off all system components (follow the Turn On Procedure in reverse)   * Turn off all system components (follow the Turn On Procedure in reverse)
  